Note: It is important that the red tow indicator of the Fractometer II 
is positioned clockwise to the zero mark. 

ENGLISH 

Instruction Manual 

For the measurement of the bending strength the 2 levers are pushed 
together as long as it takes to break the core sample; During that time 
one should make sure to focus on the pointer of the protractor where 
the bending angle of the core sample can be read of the scale.  
 
With the ring of the breaking angle measurement device the angle 
Measurement can be adjusted to zero (neutral point adjustment). 

Biegehebel 

Bruchwinkelmesser 

When measuring the radial bending moment the core sample fails first on  
the tension side. Therefore the radial bending strength can be used as a 
measurement for the radial tension strength. The result can be read off the 
tow indicator. The yellow scale on the outside indicates the mechanical 
property in Mpa (mega Pascal); the internal blue scale displays the applied 
strength in Newton Units. 
 
 

IMPORTANT NOTE:

  

The pointer that measures the bending break angle does not stop when 
the core sample breaks; Therefore, it is important to focus on the pointer 
during the bending process!

 

 

Scale for Bending 

Strength Measurement 

(left side) 

Biegehebel 

Breaking angle measurement device 

Scale for 

Compression 

Strength 

Measurement 

(right side) 

Bending lever 

Summary of Contents for FRACTOMETER II

Page 1: ...GUARANTEE POLICY IML Instrumenta Mechanik Labor System GmbH Parkstra e 33 69168 Wiesloch Germany Phone 49 0 6222 6797 0 Fax 49 0 6222 6797 10 Website www iml de E mail info iml de FRACTOMETER II made...

Page 2: ...Procedure of the measurement Fibre direction during the bending strength measurement 4 5 Measurement of the compression strength Procedure of the measurement Fibre direction during the compression str...

Page 3: ...for The bending angle display can be examined with the test pin Bending Angle in zero position Insert the test pin into the bending device and Display link up the bending lever Read off the value of...

Page 4: ...nststoffgeh use welches ber den Zuwachsbohrer geschoben wird ist auch eine Verletzungsgefahr beim Ausdr cken der Bohrkerne weitestgehend ausgeschlosse Ejector 2 INCREMENT BORER With the use of the inc...

Page 5: ...achte Kraft Diese Kraft entspricht dann der Versagenslast F ENGLISH Introduction THE STRUCTURE OF THE FRACTOMETER II The Fractometer II consists of 6 basic components Lever for the measurement of the...

Page 6: ...n axial direction to the trunk In roots and curved trunks we will see perpendicular stress or radial tensions The radial tension strength is the resistance which the trunk opposes to a failure by perp...

Page 7: ...uchwinkelmesser When measuring the radial bending moment the core sample fails first on the tension side Therefore the radial bending strength can be used as a measurement for the radial tension stren...

Page 8: ...Fractometer II is positioned counter clockwise to the zero mark Make sure that the fiber direction of the core sample is positioned parallel to the applied force that presses onto the core sample IMPO...

Page 9: ...w indicator stops with the measured value and that can be read off immediately in Mpa and or Newton units ENGLISH 7 Instruction Manual Compression Failure Compression Strength Failure Distance Point 2...

Page 10: ...are only an example Don t forget to focus on the overall structure and health of a tree Resources for Tables and Graphs Different Strengths of Green Woods G tz 2000 Research Center in Karlsruhe German...

Page 11: ...9 ENGLISH Compression Strength in Fiber Direction Tables and Graphs...

Page 12: ...he contract of purchase Please also refer to the manufacturer s General Terms of Business INSTRUCTIONS FOR UTILIZATION The user accepts the manufacture s instruction for utilization when he operates t...

Page 13: ...nclusions with respect to the examined object ex Core sample of tree will be left to the sole responsibility of the person who performs the examination The manufacturer of the instruments will not be...

Reviews: