![IMC Cansas - SCI8 Getting Started Download Page 44](http://html1.mh-extra.com/html/imc/cansas-sci8/cansas-sci8_getting-started_2045717044.webp)
44
Pin configuration and power supply
imc CANSAS Getting Started - Doc. Version 1.9 - 05.12.2014
© 2014 imc Meßsysteme GmbH
Divider-plug CAN/DSUB-U6D / U4D
-in1
+in1
-in3
+in4
-in4
+in5
+in3
+Iref
-in5
+in6
-in6
+in2
+in1
-in1
-Iref
Iref
Divider-plug: CANSAS
(CAN/DSUB-U6D / U4D)
MUX
8
9
2
10
labelling
DSUB-15
Pins
15
7
14
6
13
5
12
11
4
1
2
3
6
7
10
11
4
5
8
9
12
terminal-
numbers
-in2
+in2
-in3
+in3
-in4
+in4
-in5
+in5
-in6
+in6
-in2
3
no divider in
CANSAS ISO-8
no divider in
CANSAS ISO-8
5k
495k
2.2 nF
2.2 nF
2.2 nF
2.2 nF
2.2 nF
2.2 nF
Note
The maximum signal voltage is 60 V!
Note that the resistance dividers are not symmetrical!
The divider plug CAN/DSUB-U6D serves to connect six voltage signals to imc CANSAS-C12.
The accuracy ratings only apply to devices correctly configured and connected. Please make sure
that no inputs used may be not connected, otherwise overshoot may occur at inputs, thus
affecting other channels and resulting increased measurement error.