Page 24
Glass Plate Setup and Assembly Manual
B.2.2 White
Balance
Perform the function as follows:
1. Find the section
White Balance Adjustments
and click the button White Balance .
2. Place the
Line Reference Target LRT-BE3-x
under the glass plate as shown on the
screen. The screen shows a blinking area at the LRT-BE3-x.
3. Place the
White Reference Target 110x670 mm
at the marked position.
Picture 27: “White Balance” screen
4. Press the Next Step button.
5. After the White Balance function has finished, the results will be displayed. As long as
the displayed text is green, the result is ok. Any error will be shown in red followed by
some explanatory remarks.
6. Click the button Back to Adj. Menu in the section
Menu Controls
to return to the
adjustments menu (see Picture 25).
Or
7. Click the button Back to Main Menu in the section
Menu Controls
to return to the
“Poweruser” menu (see Picture 24).
At this point, all mandatory setup functions have been completed. Click the
Scan2Net logo on top of the screen and go to
Launch Scan Application
to start
scanning.