Setup Manual
Page 79
Place two CSTT test targets on the glass plate as shown at the screen.
Click on the Next Step button to start scanning the test targets. The test sequence will
take approximately 30 seconds.
After scanning, the image will not be displayed.
A small window – depending on the browser used for scanning – opens.
Select between opening the image with an appropriate software application and saving
the image.
The default image name contains the test target name, the device type and the serial
number of the scanner.
Picture 60 shows as example the window of the Mozilla Firefox browser.
Picture 60: Request after scanning the test target
The file name contains the following information:
CSTT-WT25-600-0019999c96fe
CSTT: Test target name.
WT25-600: Device type, here WT25-600
0019999c96fe: Serial number of the scanner.
This information is helpful for service technicians to find the scanner specific data in the
database.
The test target image contains information which allows analysis of the current settings of
the scanner and comparison with the factory settings.