5.
External Power ON/OFF
(Same as
ON/OFF
key function)
Connecting #24 (EX_Power) and #30 (GND) of the communications
port turns ON the gauge and connecting #24 (EX_Power) and
#30 (GND) while #29 (Shift) and #30 (GND) are connected
turns OFF the gauge.
6.
External Zero
(Same as
ZERO
key function)
Connecting #25 (EX_ZERO) and #30 (GND) of the communications
port resets the force display. (Tare, peak reset, or peak reset only
are dependent on the Zero/Tare Reset settings, see page 41).
7. External Send
(same as the
SEND
key function)
Connecting #26 (EX_SEND) and #30 (GND) of the communica-
tions port sends data to gauge memory (up to 1,000 data) and an
external device and holds the display while the signal is closed.
8. External Peak
(same as
PEAK
key function)
Connecting #27 (EX_PEAK) and #30 (GND) of the communications
port cycles real time, compression peak and tension peak.
page 13
Use contact closure only and
DO NOT
apply voltage
to #24 through #30 port pins.
!
4. Optional Data Acquisition Accessories
SW-1 Data Acquisition Software
Capture and analyze force data from gauges with an RS-232 port. A
log of all the data is displayed along with a graph and statistics.
ZT-R Force Recorder Software
Records and processes 2,000 data per second from ZTS gauges and
generates a force profile graph with statistics. Zoom in to examine
force vs. time values at individual data points. Select part of the graph
to detect the first peak or generate statistics for that time interval.
Align and compare multiple graphs. Save graphs or export data to CSV.
Cables
CB-108
10' Analog cable
CB-208
10' RS-232C cable, 9 pin female
CB-308
10' Digimatic Cable