![ILX Lightwave OMH-6727B User Manual Download Page 17](http://html1.mh-extra.com/html/ilx-lightwave/omh-6727b/omh-6727b_user-manual_2044778017.webp)
I N T R O D U C T I O N A N D S P E C I F I C A T I O N S
Specifications
4_03
OMH-6727B
3
C H A P T E R
1
Specifications
Input Specifications
Power Measurement
Range
100 nW ~ 1 W
2
; -40 dBm ~ + 30 dBm
2
Accuracy
+3.5% (950 to 1500 nm)
1,3
Wavelength Measurement
Range
950 to 1650 nm
4
Accuracy
+ 1.0 nm
Temperature coefficient
< + 0.03 nm /
o
C
Linearity with power change
< + 0.5 nm
(Maximum change in measured wavelength due to power change)
Sampling Rate
Slow
16 samples @ 60 ms / sample
Medium
4 samples @ 60 ms / sample
Fast
1 sample @ 60 ms / sample
Input Bandwidth
Slow
1 Hz
Medium
10 Hz
Fast
60 Hz
Analog Output Specifications
Analog output representing the power measurement is available at the rear panel of the multimeter.
Refer to the OMM-6810B Instruction Manual for details.
Display
Left Display (Power)
Units
Linear Power
Log Power
pW, nW,
µ
W, mW, W,
∆
P
dBm, dB
Range
Linear Power
Log Power
0.000 nW to 999.999 W
-99.999 to 99.999 dBm / dB
Resolution (Slow Update Rate)
Linear Power
Log Power
0.001 pW
0.001 dB
Resolution (Medium / Fast Update Rate)
Linear Power
Log Power
0.01 pW
0.01 dB