Action
Performs internal self−test, then reports results.
Response
0 = test completed with no errors.
Non−zero = test not completed, or was completed with errors.
Example
TST?
Action
Prevents the FPM−8200 from executing any further commands until OPC (operation complete) sta−
tus is true.
Note
Operation complete is defined for the FPM−8200 as:
The controller is idle.
Internal non−volatile memory write cycles are completed.
A power measurement result is available.
No calibration processes are running.
This command can be used to make the FPM−8200 wait until an operation is complete before con−
tinuing.
Example
*WAI
—
Wait until OPC status is true.
Action
Sets the wavelength (in nanometers) to be used in calibrating detector response.
Value range
800 to 1600, in increments of 1
Notes
The FPM−8200’s InGaAs detector produces current in proportion to light input. The amount of cur−
rent varies also with wavelength. For this reason, it is important to give the FPM−8200 correct
wavelength information.
Calibration points are every 10 nm. When you give the FPM−8200 a wavelength between two cali−
bration points, the resulting calibration factor is a linear interpolation between the two points.
Example
WAVE 1552˙Set wavelength response to 1552 nm. The calibration factor will be interpolated to
include 20% of the change in calibration points from 1550 nm to 1560 nm.
Action
Requests the wavelength to be used in calibrating detector response.
Response
A GPIB−standard nrf value of the current wavelength setting.
See WAVE above.
Note
This information is available on the FPM−8200 front panel.
Example
WAVE?
—
Response 1552 means FPM−8200 wavelength response is set to 1552 nm. The
calibration factor will be interpolated to include 20% of the change in calibration points from 1550
nm to 1560 nm.
*TST?
Test?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
*WAI
Wait
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
WAVE <nrf value>
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
WAVE?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
Summary of Contents for FPM-8200
Page 2: ......
Page 8: ...L I S T O F FIG U R E S vi FPM 8200...
Page 10: ...L I S T O F TA B L E S viii FPM 8200...
Page 16: ...WA R R A N T Y xiv FPM 8200...
Page 36: ...FR O NT PA N E L O P E R A T I O N Operating Guidelines 20 FPM 8200 C H A P T E R 2...
Page 96: ...80 FPM 8200 A P P E N D I X A...