![ILX Lightwave FPM-8200 User Manual Download Page 63](http://html1.mh-extra.com/html/ilx-lightwave/fpm-8200/fpm-8200_user-manual_2044775063.webp)
G P I B C O M M A N D R E F E R E N C E
GPIB Command Reference
01_10
FPM-8200
47
C H A P T E R
4
Action
Requests the value in the condition status enable register.
Response
Notes
Enabled/disabled conditions can be set by ENABle:COND.
Changing condition status may be monitored by COND?.
See Chapter Three for more information about register structure.
Examples
ENAB:COND?˙Response 24576 means that hardware or software error conditions will be
summarized and reported in the status byte bit 3 (8192 + 16384 = 24576).
Enable:cond
˙
Response #H6000 is the same as 24576, except using hexadecimal numbering. See
RADix on page 57.
ENABle:COND?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
Over-range
Numeric
Value:
Under-range
Ready for calibration data
Calculation error
Hardware error
Software error
= Unused bits
4
8
2048
4096
8192
16348
Summary of Contents for FPM-8200