G P I B C O M M A N D R E F E R E N C E
GPIB Command Reference
52
FPM-8200
C H A P T E R
4
Action
Requests the value in the event status register.
Response
Notes
Enabled/disabled events can be set by ENABle:EVEnt.
Status bits are set by events, and cleared when read or cleared by command. See *CLS and
EVEnt? commands.
The measurement ready bit is set whenever the display value is updated. For example, in slow
mode a new measurement is ready every 5 seconds. See FILTer below.
See Chapter Three for more information about register structure.
Examples
EVE?˙Response 2052 means that over−range and measurement ready events occurred since the
last EVE? inquiry.
Event?
˙Response #H804 is the same as 2052, except using hexadecimal numbering. See RADix
on page 57.
Action
Selects measurement averaging and display update rate.
Notes
Measurement samples are taken every 50 mSec.
This command does not affect the bar graph update rate.
Examples
FILTER SLOW
—
Updates the display every 5 seconds with the average of 100 measurements.
Filter med
—
Updates the display every 0.5 seconds with the average of 10 measurements.
FILT Fast
—
Turns off measurement averaging. Updates the display every 0.05 seconds.
EVEnt?
Event status?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
FILTer
SLOW
MED
FAST
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
Over-range
Numeric
Value:
Under-range
Ready for calibration data
Calculation error
Hardware error
Software error
= Unused bits
4
8
2048
4096
8192
16348