Cleaning & Inspection
Swift F1
Quick Reference Guide
•
Brush the surface of the V-grooves & Stripper and clean
with a cotton swab moistened with alcohol.
•
Remove the remaining alcohol in the V-grooves with a dry
clean cotton swab.
•
Use the cleaved optical fiber to remove contaminants
from the V-grooves if any remain.
•
Clean the surface of the wind protector prisms with a lens
Clean the V-Groove
Clean the Stripper
•
Turn OFF the F1 before Cleaning & Inspection.
•
Be careful handling the ELECTRODES.
•
Brush the V-Groove & Stripper frequently before / after using F1.
•
Be careful handling STRIPPER BLADE PART which is very sensitive.
Clean the surface of the wind protector prisms with a lens
cleaner or cotton swab.
•
Remove the electrodes before cleaning the object lenses.
•
Clean the surface of the lens(X axis & Y axis) softly with
a lens cleaner or cotton swab.
•
Remove remaining alcohol with a dry clean cotton swab.
•
Turn on the power and check the clarity of the lens into
the display LCD monitor.
Press the “▽” cursor button to change the screen and inspect
the surface of the lens.
Wind Protector Prism
Clean the Object Lens
Routine Maintenance
Replacing Electrodes
•
Turn OFF the F1 before doing process.
•
Be careful handling the ELECTRODES.
•
Brush the V-Groove &Stripper frequently before / after using F1.
•
Cleaning Electrodes once every 500 times arc with a cotton swab
moistened with alcohol.
1
F
th
i
l t AUX
Battery Maintenance
•
Discharge the battery completely
at least once every 3 months in
order to avoid memory effect.
•
Storage temperature should be
between -20℃ and 30℃.
•
Recharge the battery every 6
p
g
1.
From the main menu, select AUX menu, press
ENTER, REPLACE ELECTRODES, and press ENTER.
2.
Loosen the thumb-screws and remove the electrodes.
3.
Clean the new electrodes and install.
4.
From the main menu, perform the ARC Calibration.
Select AUX menu, perform the Stabilize Electrode
in order to stabilize electrodes.
5.
From the main menu, select AUX menu, set the “O”
arc count through the Clear ARC Count
•
Recharge the battery every 6
months when not in use.
Routine Diagnostics
The following tests are found in
the OTHER menu in the system
menu, and should be performed
once every 3 months.
•
Dust Check
•
Motor Drive
•
LED Check
Contact :
Contact :
/ +82
/ +82--42
42--671
671--5609~11
5609~11
arc count through the Clear ARC Count
LED Check