Error  Messages

Swift F1

Quick Reference Guide

Symptom

Cause

Solution

Clad 

Deviation

Dust in V-Groove or 

the chip of core wire 

guide.

Cleaning V-Groove

or the chip of core 

wire guide.

Clad Angle

Dust in V-Groove or 

the chip of core wire 

guide.

Cleaning V-Groove

or the chip of core 

wire guide.

Symptom

Cause

Solution

Bubble

Bad condition of the 

cross section of 

fiber.

Checking whether 

fiber cleaver is 

working properly.

Too low amount or 

short time of initial 

arc discharge.

Increase the amount 

and/or time of initial 

are discharge.

g

g

g

Bad condition of the 

cross section of fiber.

Checking whether

fiber cleaver is 

working properly.

Clad

Deformation

Bad condition of the 

cross section of fiber.

Checking whether 

fiber cleaver is 

working properly.

Too low amount or 

short time of initial 

arc discharge

Increasing the 

amount and time of 

initial arc discharge.

Ch ki

h th

Separation

Too high amount or 

long time of initial 

arc discharge.

Increase overlapping.

Decreasing the 

amount and/or time 

of initial arc 

discharge.

Thick

Too long fiber 

length.

Reducing the length 

of duplicated area.

Clad Flexion

Bad condition of the 

cross section of fiber.

Checking whether 

fiber cleaver is 

working properly.

Too low amount or 

short time of initial 

arc discharge.

Increase the amount 

and time of initial 

arc discharge.

Burning

Bad condition of the 

cross  section of 

fiber.

Checking whether 

fiber cleaver is 

working properly.

Dust still remaining 

after cleaning the

Cleaning carefully

the fiber or

Thin

Improper discharge 

amount.

Conducting

discharge amount 

calibration

A few improper 

discharge factors.

Adjusting initial 

discharge amount, 

initial discharge time 

and overlapped area.

Line

Several number of 

improper discharge 

Adjusting initial 

discharge amount, 

initial discharge time

after cleaning the 

fiber or conducting 

cleaning discharge.

the fiber or 

increasing the arc 

discharge time.

p p

g

factors.

initial discharge time 

and overlapped area.

Contact  ILSINTECH

Add:

#703 Gwanpyeong-dong  Yusung-gu

Daejeon 305-509 Korea

Contact : 

Contact : 

[email protected]

[email protected]

/   +82

/   +82--42

42--671

671--5609~11

5609~11

Daejeon 305 509, Korea

Tel:

+82-42-671-5609~11

Fax:

+82-42-671-5612

E-mail:

[email protected]

www.ilsintech.com

Summary of Contents for SWIFT F1

Page 1: ...1 2 Charger AC DC Adaptor AC Code Battery Capacity Check Remaining battery capacity is displayed on both the top right corner of the Monitor and the AC power supply for the specified charger is AC 100...

Page 2: ...otector Prism Clean the Object Lens Routine Maintenance Replacing Electrodes Turn OFF the F1 before doing process Be careful handling the ELECTRODES Brush the V Groove Stripper frequently before after...

Page 3: ...e because the height directly effect SPLICE LOSS 1 2 3 Adjustment of Blade Height 2 1 1 Loosen the set screw at the side of the cleaver detached from the F1 a little bit approx 2 turns with a wrench 2...

Page 4: ...into the proper fiber holder 250 900 2 5 and Ferrule type holders Sleeve Tube Fiber Ferrule Ferrule Cap yp 4 Stripping Operation Warm up stripper heater First close holder part cover after mounting th...

Page 5: ...MUST BE PLACED ON CORRECT POSITIONS AND BE CAREFUL THE FIBER FROM ANY DUST OR CONTAMINANT Dispatch fibers from the holders and gently move the sleeve tube to the spliced part of fiber For ferrule soft...

Page 6: ...Indoor Cable Lever Block Heater Block L Heater Block R Type Description Type Description 0 25 0 9mm Fiber 0 25 0 9mm Fiber 2 0 2 4mm Cable LC Connector 2 0 2 4mm Cable 3 0mm Cable 3 0mm Cable 3 0mm C...

Page 7: ...n the Set Screw on Heater Block L 2 Mount 1 Lever Block Tighten the Set Screw 3 Loosen the Screw on Bottom of Heater Block R 4 Replace the Heater Block R for SC FC Connector Tighten the Screw 5 Placed...

Page 8: ...Loosen the Set Screw on Heater Block L 2 Mount 3 Lever Block Tighten the Set Screw 3 Loosen the Screw on Bottom of Heater Block R 4 Replace the Heater Block R for SC FC Connector Tighten the Screw 5...

Page 9: ...pe 2 0mm Cable PVC Outer Sheath Cut Kevlar 5 Insert Boot Heat Shrinkable Tube Stopper Spring and Sleeve into the 2 0mm Cable Place 2 0mm Cable on HF 2 5 Holder Sleeve use environmental condition Temp...

Page 10: ...ing Sleeve 8 Placed on Sleeve on Heater as PIC 9 Splicing Stopper Assembly APC Connector Red Mark in the APC Ferrule must head to the upward then assembly to the Plug Frame 10 Sleeve 2 Heating Heat Sh...

Page 11: ...erence Guide Step Description Procedure 1 Place 2 0 3 0mm Cable on Holder Cutting Kevlar point 2 Cut Kevlar Approx 20mm 3 Heat Shrink the Sleeve Sleeve Kevlar inside of Sleeve Contact Contact sales il...

Page 12: ...ly Process Swift F1 Quick Reference Guide Step Description Procedure 4 Assemble Boot 5 Assemble Housing Clamp Cap Boot Clamp Cap 6 Complete Housing Contact Contact sales ilsintech com sales ilsintech...

Page 13: ...arc discharge Thick Too long fiber length Reducing the length of duplicated area Clad Flexion Bad condition of the cross section of fiber Checking whether fiber cleaver is working properly Too low amo...

Reviews: