Maintenance and Servicing 

6.2.3 Test 

Connect a vacuum measuring device to the intake port and check the ultimate pressure. If 
the device is working properly, then the figure stated in the technical data must be attained 
within a maximum of one minute. 
 

6.3 

Maintenance of the diaphragm pump 

• 

see the attached "Diaphragm pumps" operating manual 

 

6.4 

Maintenance of other components 

• 

Check that the glass drip pan is undamaged, cleaned and if necessary replace.  

• 

Empty the drip pan in a timely manner 
(Observe all disposal specifications as applicable to hazardous substances!) 

• 

Check screw connections for tightness and tightened when necessary.  

• 

Check vacuum hoses for leaks and, if necessary, replace them. 

• 

Check for leaks and, if necessary, replace the formed gasket located at the head piece of 
the drip pan 

• 

Change the membrane filter at regular, defined intervals. 

• 

Test performance of the quick-fitting couplings with shut-off valves 

 

 

WARNING ! 

Only perform the work that is described here, and that which is permitted to be done 
by the user. 
All other maintenance and service work may only be performed by the manufacturer 
or a dealer authorized by him. 
Beware of the device parts being possibly contaminated by hazardous substances. 
Please observe the applicable disposal regulations ! 
Wear protective clothing if there is contamination.  

 
 

6.5 Damage 

Report 

You find the form of the damage report to the Download on our web page 

http://www.ilmvac.de

 and/or 

http://www.ilmvac.com

 in the menu "service" and "Downloads". 

If you should not have an entrance to the Internet, you can request the form also gladly with 
us, company Ilmvac GmbH. 

 

 

WARNING ! 

Incomplete or incorrectly completed damage reports may endanger the service per-
sonnel! 
Give full information in the damage report, in particular regarding a possible con-
taminating. 

 

 112012 

15 

Summary of Contents for biovac 106

Page 1: ...Operation Manual Safety Aspiration System Type biovac 106 112012 2008 04 04...

Page 2: ...owed without the prior written permission of ILMVAC GmbH All rights under the copyright laws are expressly reserved by ILMVAC GmbH We reserve the right to make changes and amendments ILMVAC GmbH Am Vo...

Page 3: ...Accessories 9 4 Technical Data 10 4 1 Intake Pressure Pumping Speed Diagram 10 5 Installation and Operation 11 5 1 Unpacking 11 5 2 Installation and Connection 11 5 2 1 Schematic representation of the...

Page 4: ...ion System must have the technical competence required to perform the work that has to be done The user must authorize the operating personnel to do the work that has to be done The personnel must hav...

Page 5: ...tions EN 61010 1 Safety for laboratory devices EN 50081 1 2 Electromagnetic compatibility EMC Generic specification Interference resistance for residential business and industrial areas and small busi...

Page 6: ...ase of portable devices The local regulations apply in foreign countries Please note the following when connecting to the electrical power supply system The electrical power supply system must have a...

Page 7: ...gases Prevent such substances from leaking out of the appliance or pump Treat these substances according to the applicable environmental protection regulations The diaphragm pump control valves and h...

Page 8: ...ket The connections to the diaphragm pump are established by using quick fitting couplings 4 which are provided with a two side stop valve each on both sides The chemical resistant diaphragm pump used...

Page 9: ...s soon as the maximum level is reached 3 3 Areas of Application The ILMVAC biovac 106 Safety Aspiration System is intended for the safe and precise extraction of liquid and non explosive excess fluids...

Page 10: ...ature C 40 Bearing maintenance free Reference surface sound pressure level DIN 45635 part 13 dB A 45 Voltage AC asynchronous motor V Hz 115 230 Frequency Hz 50 60 Power W 0 45 Fine fuse A F 3 15 Opera...

Page 11: ...claim under warranty the device must be returned in packaging that is suitable for protecting it during transport 5 2 Installation and Connection 1 Set the safety aspiration system on a flat and horiz...

Page 12: ...ower supply either 115 or 220 V 10 Set mode selector switch to Hand or Foot The safety aspiration system is switched on and off at the operating switch illuminated rocker switch Press the foot switch...

Page 13: ...tection may be used 5 5 Scrap Disposal CAUTION The safety aspiration systems must be disposed of in accordance with the 2002 96 EC guideline and the specific national regulations Contaminated pump sy...

Page 14: ...materials and clean the pump thoroughly before handing it over to third parties The operator may perform maintenance work to the extent indicated below 6 2 Maintenance of the aspiration system 6 2 1...

Page 15: ...he membrane filter at regular defined intervals Test performance of the quick fitting couplings with shut off valves WARNING Only perform the work that is described here and that which is permitted to...

Page 16: ...Connected apparatus leaks connecting elements leak Identify and seal the leak replace the seals and hoses if necessary Diaphragm pump leaks pipes or hoses Check the hose connections between the pump h...

Page 17: ...quantity serial number and order number CAUTION Ilmvac is not liable for any damage caused by the installation of any parts not sup plied by the manufacturer 8 1 Maintenance kit Designation Order no...

Page 18: ...Spare Parts Overview 8 2 Spare parts view Fig 5 Exploded view biovac 106 18 112012...

Page 19: ...ocker switch 16 4 A green 1 825186 1 29 Voltage voter 115 230 V 1 825187 30 Socket for non heating apparatus 1 825274 5 31 Casing foot 5 829112 32 Rubber plate 1 112547 02 33 Guard ring 1 112561 34 Co...

Reviews: