Maintenance and Servicing
6.2.3 Test
Connect a vacuum measuring device to the intake port and check the ultimate pressure. If
the device is working properly, then the figure stated in the technical data must be attained
within a maximum of one minute.
6.3
Maintenance of the diaphragm pump
•
see the attached "Diaphragm pumps" operating manual
6.4
Maintenance of other components
•
Check that the glass drip pan is undamaged, cleaned and if necessary replace.
•
Empty the drip pan in a timely manner
(Observe all disposal specifications as applicable to hazardous substances!)
•
Check screw connections for tightness and tightened when necessary.
•
Check vacuum hoses for leaks and, if necessary, replace them.
•
Check for leaks and, if necessary, replace the formed gasket located at the head piece of
the drip pan
•
Change the membrane filter at regular, defined intervals.
•
Test performance of the quick-fitting couplings with shut-off valves
WARNING !
Only perform the work that is described here, and that which is permitted to be done
by the user.
All other maintenance and service work may only be performed by the manufacturer
or a dealer authorized by him.
Beware of the device parts being possibly contaminated by hazardous substances.
Please observe the applicable disposal regulations !
Wear protective clothing if there is contamination.
6.5 Damage
Report
You find the form of the damage report to the Download on our web page
http://www.ilmvac.de
and/or
http://www.ilmvac.com
in the menu "service" and "Downloads".
If you should not have an entrance to the Internet, you can request the form also gladly with
us, company Ilmvac GmbH.
WARNING !
Incomplete or incorrectly completed damage reports may endanger the service per-
sonnel!
Give full information in the damage report, in particular regarding a possible con-
taminating.
112012
15