3 - Technical Data
System Modules
Infinium LiHa and Infinium RoMa Operating Manual - Document # 1000000110155 v00
3 - 11
Liquid Level
Detection
Each tip can individually detect the surface of a conductive liquid by measurement
of changes in capacitance. Each channel has an individual liquid detection.
Generally, detection of conductive liquids of following volumes is possible:
50 µl: low-conductive liquid in microplates with round bottoms
100 µl: conductive liquid in sample tubes with a diameter of 10 or 13 mm
150 µl: conductive liquid in sample tubes with a diameter of 16 mm
5 ml: conductive liquid in reagent trough
Wetted
Materials
The standard liquid system components that come into contact with either system
or sample liquid are of the following materials:
Also refer to section
Tab. 3-10
Liquid system components: materials
Component
Material
Pipetting tubing
FEP
Tubing (waste, part of aspirating tubing)
Silicone
Distributor 1:4
POM
Aspirating tubing
PVC
Wash stations, Y-connectors
PP
FaWa
FFPM (membrane), PP (body)
Pressure relief valve
PP
Valves (diluters)
PCTFE (Kel-F)
Syringes
Borosilicate glass
Syringes, seals
PTFE
Tips
Stainless steel, PTFE
a)
a) Coating
Liquid containers
HD-PE