background image

20

Operator panel and display

/// Operation

rpm

A

C

B

A

Standby key

Device switches on / enters standby  
(no disconnection from the power supply network).

B

Display 

Display the speed value.

C

Control knob 
(rotating / pressing)

Set the speed by rotating.
Start / stop the stirrer function by pressing.

21

Assembly 

/// Assembling the extension arm to the overhead 
stirrer

Stirrer

Extension arm

Hexagonal socket screw

Hexagon socket 

offset screw key

Ensure that the extension arm is fitted securely.
Vibration may cause the screw to become loose. It is therefore necessary for 
safe use to regularly check that the extension arm is attached securely. Tighten 
the hexangular bolt as required.

Summary of Contents for MINISTAR digital

Page 1: ...MINISTAR digital...

Page 2: ...9 8 Device setup MINISTAR digital 1 Operator panel and display 2 Chuck 3 Extension arm 4 Power socket 5 USB interface 6 Stirring shaft cap What you should watch out for Attention 6 5 3 2 4 1...

Page 3: ...ns Keep the operating instructions in a place where it can be accessed by everyone Ensure that only trained staff work with the device Follow the safety instructions guidelines occupational health and...

Page 4: ...anger Do not use the device in explosive atmospheres it is not EX protected With substances capable of forming an explosive mixture appropriate safety measures must be applied e g working under a fume...

Page 5: ...duce the speed if the medium splashes out of the vessel because the speed is too high the device is not running smoothly the device begins to move around because of dynamic forces an error occurs Acce...

Page 6: ...ntrol Warranty MINISTAR 20 Intended use Basics Use For mixing stirring liquids from low to medium viscosities with various stirring tools Stand device chuck points down Range of use indoor use only La...

Page 7: ...oad range for instance if the viscosity has increased at a particular stage of the process the speed will reduce to a degree sufficient to keep the torque on the stirrer shaft within the rated torque...

Page 8: ...rotating Start stop the stirrer function by pressing 21 Assembly Assembling the extension arm to the overhead stirrer Stirrer Extension arm Hexagonal socket screw Hexagon socket offset screw key Ensur...

Page 9: ...djusted when the equipment is stationary and the power supply is disconnected Assembling the stirring element using the chuck Chuck Stirring element Chuck key Warning The stirring element must only be...

Page 10: ...erial water containing tenside isopropyl alcohol Cosmetics water containing tenside isopropyl alcohol Foodstuffs water containing tenside Fuel water containing tenside Dirt Cleaning agent For material...

Page 11: ...d sensor fault or overload Effect motor off Solution disconnect the device from the power supply If the actions described fails to resolve the fault or another error code is displayed then take one of...

Page 12: ...essories An overview General accessories R 1825 Plate stand R 1826 Plate stand R 1827 Plate stand R 182 Cross sleeve RH 3 Strap clamp R 300 Stirring shaft protection with 2 screws FK 1 Flexible coupli...

Page 13: ...e USB Permitted on time 100 Nominal voltage 24 VDC Max current 2900 mA Max input power 69 W Max output power at stirring shaft 42 W IP code according to EN 60529 IP 54 Contamination level 2 Protection...