background image

 13

12

/// Device setup 

Danger! (   )

Users must make sure that the standby key of the IKA device can be accessed 
immediately, directly and without risk at any time. If installation or positioning 
cannot ensure this access at all times, an additional 

EMERGENCY STOP

 switch 

that can be easily accessed must be installed in the work area.

Warning! (  )

Ensure stable assembly. The vessel used for stirring must be secured.
Use stirring shaft protective equipment!

 

Set up the stand in a spacious area on an even, stable, clean, non-
slip, dry and fireproof surface.

 

Remove the chuck key from the chuck before turning on the device.

 

› Place the power supply outside the working area of the device. 

 

› Avoid knocking and impacting on the lower end of the shaft and the chuck 

gear teeth. Even minor, invisible damage can lead to imbalance and uneven 
shaft action.

 

› All screw connections must be properly tightened.

 

› The high torque developed by the device requires particular care in the choice 

of stand, cross sleeve and anti-rotation element for the agitating vessel.

/// Working with the device

Danger! (   )

 

› Do not use the device in explosive atmospheres, it is not EX-protected. 

 

› With substances capable of forming an explosive mixture, appropriate safety 

measures must be applied, e.g. working under a fume hood.

 

› To avoid body injury and property damage, observe the relevant safety and 

accident prevention measures when processing hazardous materials.

 

› The device is not suitable for hand-held operation.

 

› Do not touch rotating parts during operation!

Warning! (  )

Pathogenic materials must only be processed in closed vessels under a suitable 
fume hood. If you have any questions, please contact IKA application support.

The operation of a free rotating shaft end is prohibited. Therefore, for safety reason, 
only insert through the stirring tool over the upper edge of housing at standstill.

Caution! (   )

Make certain that the unit is set at the lowest speed before commissioning; oth-
erwise, the unit will begin running at the speed set in last operation. Gradually in-
crease the speed. Please observe the max. permitted speed for the stirring element. 
Never set higher speed.

Covers or parts that can be removed from the device without tools must 
later be refitted to ensure safe operation. This will prevent the infiltration 
of foreign objects, liquids and other contaminants.

Summary of Contents for MINISTAR digital

Page 1: ...MINISTAR digital...

Page 2: ...9 8 Device setup MINISTAR digital 1 Operator panel and display 2 Chuck 3 Extension arm 4 Power socket 5 USB interface 6 Stirring shaft cap What you should watch out for Attention 6 5 3 2 4 1...

Page 3: ...ns Keep the operating instructions in a place where it can be accessed by everyone Ensure that only trained staff work with the device Follow the safety instructions guidelines occupational health and...

Page 4: ...anger Do not use the device in explosive atmospheres it is not EX protected With substances capable of forming an explosive mixture appropriate safety measures must be applied e g working under a fume...

Page 5: ...duce the speed if the medium splashes out of the vessel because the speed is too high the device is not running smoothly the device begins to move around because of dynamic forces an error occurs Acce...

Page 6: ...ntrol Warranty MINISTAR 20 Intended use Basics Use For mixing stirring liquids from low to medium viscosities with various stirring tools Stand device chuck points down Range of use indoor use only La...

Page 7: ...oad range for instance if the viscosity has increased at a particular stage of the process the speed will reduce to a degree sufficient to keep the torque on the stirrer shaft within the rated torque...

Page 8: ...rotating Start stop the stirrer function by pressing 21 Assembly Assembling the extension arm to the overhead stirrer Stirrer Extension arm Hexagonal socket screw Hexagon socket offset screw key Ensur...

Page 9: ...djusted when the equipment is stationary and the power supply is disconnected Assembling the stirring element using the chuck Chuck Stirring element Chuck key Warning The stirring element must only be...

Page 10: ...erial water containing tenside isopropyl alcohol Cosmetics water containing tenside isopropyl alcohol Foodstuffs water containing tenside Fuel water containing tenside Dirt Cleaning agent For material...

Page 11: ...d sensor fault or overload Effect motor off Solution disconnect the device from the power supply If the actions described fails to resolve the fault or another error code is displayed then take one of...

Page 12: ...essories An overview General accessories R 1825 Plate stand R 1826 Plate stand R 1827 Plate stand R 182 Cross sleeve RH 3 Strap clamp R 300 Stirring shaft protection with 2 screws FK 1 Flexible coupli...

Page 13: ...e USB Permitted on time 100 Nominal voltage 24 VDC Max current 2900 mA Max input power 69 W Max output power at stirring shaft 42 W IP code according to EN 60529 IP 54 Contamination level 2 Protection...

Reviews: