Maintenance and cleaning

The device is maintenance-free� It is only subject to the natural wear and tear of components and 
their statistical failure rate�

/// Cleaning

For cleaning disconnect the mains plug!
Use only cleaning agents which have been approved by IKA to clean the devices:
These are: water (containing surfactant) and isopropyl alcohol�

›  Wear protective gloves during cleaning the devices�
›  Electrical devices may not be placed in the cleansing agent for the purpose of cleaning�
›  Do not allow moisture to get into the device when cleaning�
›  Before using another than the recommended method for cleaning or decontamination, the 

user must ascertain with  that this method does not destroy the device�

/// Repairs

Please send in device for repair only after it has been cleaned and is free from any 

materials which may constitute a health hazard.

For repair, please request the “Decontamination Certificate” from 

IKA

, or download printout 

of it from the 

IKA

 website 

www.ika.com

If you require servicing, return the device in its original packaging� Storage packaging is not 
sufficient� Please also use suitable transport packaging�

/// Ordering spare parts

When ordering spare parts, please indicate:
›  Device type�
›  Serial number, see type plate
›  Item and designation of the spare part, see:

 www.ika.com

, spare parts diagram and spare 

parts list�

›  Software version (Briefly visible in the display when the device is switched on)�

If the action described fails to resolve the fault or another error code is displayed then take 
one of the following steps: 
›  Contact the service department 
›  Send the device for repair, including a short description of the fault� 

Error 49 - Motor hall sensor fault

Error message

›  Motor hall sensor signal lost

Effect

›  Mixing function off
›  Tempering function off

Solutions

›  Switch off and restart the device�
›  Send device for repair if the error message appear again�

Accessories

For accessiores see 

www.ika.com

Exchangable attachments

Tubes / Plates 

Max. Drehzahl 

Max. 

Temperature

Min. 

Temperature

IKA MATRIX AUFSATZ 0�5 ML

24 x 0�5 ml

 

Gefäße

2000 rpm

100 °C

RT - 15 °C 

IKA MATRIX AUFSATZ 1�5 ML

24

 

x 1�5 ml Gefäße

2000 rpm

100 °C

RT - 15 °C 

IKA MATRIX AUFSATZ 12 MM

24

  

x 11 mm bis 11�9 mm 

Gefäße 

2000 rpm

100 °C 

RT - 15 °C 

IKA MATRIX AUFSATZ CRYO

24

 

x 1�5 ml oder 2�0 ml 

Cryo-Gefäße

2000 rpm

100 °C

RT - 15 °C

IKA MATRIX AUFSATZ 2�0 ML

24

 

x 2�0 ml Gefäße

2000 rpm

100 °C

RT - 15 °C 

IKA MATRIX AUFSATZ 5�0 ML

8

 

x 5�0 ml Gefäße

1000 rpm

100 °C

RT - 15 °C 

IKA MATRIX AUFSATZ 15 ML

8

 

x 15 ml Gefäße

1000 rpm

100 °C

RT - 15 °C 

IKA MATRIX AUFSATZ 50 ML

4

 

x 50 ml Gefäße

1000 rpm

100 °C

RT - 15 °C 

IKA MATRIX AUFSATZ PCR 96

96

 

x 0�2 ml PCR-Gefäße 

1

 

x 96-Well-PCR-Platten

2000 rpm

100 °C

RT - 15 °C 

IKA MATRIX AUFSATZ PCR 384

1

 

x 384-Well-PCR-Platten

3000 rpm

100 °C

RT - 15 °C 

IKA CARRIER

1 x MTP 

1 x DWP 

3000 rpm
2000 rpm

100 °C

RT - 15 °C 

IKA Abdeckung

For IKA PCR 96, IKA PCR 384, IKA 0�5 ml, IKA 1�5 ml, IKA 2�0 ml and IKA CARRIER (MTP & DWP)�

Note:

 For even tempering and to protect the samples from unwanted exposure to light�

IKA THERMOCOVER

For IKA PCR 96, IKA PCR 384, IKA 0�5 ml, IKA 1�5 ml, IKA 2�0 ml  and IKA CARRIER (MTP & DWP)�

Note:

 Prevent the formation of condensate on vessel walls or vessel lids�

Summary of Contents for MATRIX ORBITAL DELTA PLUS

Page 1: ...IKA MATRIX ORBITAL DELTA PLUS...

Page 2: ...2 3 ENGLISH...

Page 3: ...rning symbols 6 Safety instructions 7 Unpacking 10 Operation 17 Intended use 9 Technical Data 38 Interfaces and outputs 31 Warranty 39 Maintenance and cleaning 37 Accessories 37 Operator panel and dis...

Page 4: ...arting up and follow the safety instructions Keep the operating instructions in a place where it can be accessed by everyone Ensure that only trained staff work with the device Follow the safety instr...

Page 5: ...foreign objects liquids and other contaminants Never use tubes made of glass or other fragile material Glass tubes can be smash Only use sealed micro test tube and sealed microplate deepwell plate Th...

Page 6: ...empering function E Timer button Enter the timer menu F Programs Prog button Enter the programs menu G Menu button Touch it once main menu is displayed Touch it a second time back to the working scree...

Page 7: ...temperature reached see Tempering under Timer Timer Counter symbol Mixing Indicate the Timer Counter for mixing Mixing symbol Indicate the mixing function is activated 00 10 Mixing in Intermittent mod...

Page 8: ...14 15 Inserting the deepwell plate DWP microplate MTP Inserting the micro test tubes Attaching the exchangeable attachments 1 2 1 2 1 Click 3 2 A B C D E F G H 1 2...

Page 9: ...een established with the device by pressing down the button the heating function can prevent the formation of condensate Operation Switching on IKA MATRIX ORBITAL Delta Version 1 02 003 1 01 021 Shake...

Page 10: ...tual rpm 25 1 0 300 0 00 00 0 00 01 38 0 LIMIT 1 0 LIMIT 100 0 0 00 00 99 59 59 25 0 38 0 max Setting the speed and starting mixing A IKA MATRIX 2 0 mL Fast actual C Set C Set rpm actual rpm 38 0 38 0...

Page 11: ...s 3 3 3 3 3 3 F F F F s s s s s s s s 2 2 2 Restarting the tempering and mixing function Note During the pause status the tempering mixing function can be switched off separately by touching temperin...

Page 12: ...0 mL Fast actual C Set C Set rpm actual rpm 38 0 38 0 500 500 0 00 39 0 00 50 A IKA MATRIX 2 0 mL Fast actual C Set C Set rpm actual rpm 38 0 500 0 00 40 0 00 51 38 0 500 Unlocking the buttons 1s A I...

Page 13: ...0 Fast 0 00 00 10 00 00 0 500 Quick mix Note When the tempering function is activated the Quick mix function is not available Press the Menu button G Select right menu item by touching the speed butt...

Page 14: ...ed and CW time Stop time is set Note CW time and Stop time can be set separately CCW 0 CW 300 rpm time 300 CCW Intermittent mode is activated and CCW time Stop time is set Note CCW time and Stop time...

Page 15: ...atically executed In control mode x x K The hysteresis tolerance of the actual temperature to target temperature is set e g 0 1 K The target values and settings of the segment are valid until the actu...

Page 16: ...t C or F for the temperature A check mark indicates the unit that is set for the system Display In the Background menu option you select black or white color for the working screen In the Firmware Upd...

Page 17: ...ed to the WD safety temperature limit value The error message Error 2 is displayed The above details correspond as far as possible to the recommendations of the NAMUR working party NAMUR recommendatio...

Page 18: ...ons between device and external devices PC 1 1 Cable This cable is required to connect RS 232 port to a PC 1 2 RxD 3 TxD 4 5 GND 6 7 RTS 8 CTS 9 1 RxD 2 TxD 3 4 GND 5 6 RTS 7 CTS 8 9 PC 1 2 3 4 5 6 7...

Page 19: ...r error code is displayed then take one of the following steps Contact the service department Send the device for repair including a short description of the fault Error 49 Motor hall sensor fault Err...

Page 20: ...e delivery invoice and giving reasons for the claim You will be liable for freight costs The warranty does not cover worn out parts nor does it apply to faults resulting from improper use insufficient...

Page 21: ...sales england ika com MALAYSIA IKA Works Asia Sdn Bhd Phone 60 3 6099 5666 eMail sales lab ika my CHINA IKA Works Guangzhou Phone 86 20 8222 6771 eMail info ika cn POLAND IKA Poland Sp z o o Phone 48...

Reviews: