/// Menu details

  

Tempering:

Control rate:
In the "Control rate" menu option, you can select the desired control rate between "Fast", "Mod-
erate" and "Slow"� A check mark (√) indicates the control rate is activated�

Fast: 

Fast tempering with a certain level overshooting�

Moderate: 

Moderate tempering with minimal overshooting�

Slow: 

Precise tempering without overshooting�

  

Mixing:

Intermittent mode:
In the "Activate" menu option, you can activate or deactivate the "Intermittent mode"� A check 
mark (√) indicates that the function is activated�
Once the "Intermittent mode" is activated, the set values for "CW time", "Stop time" and 
"CCW time" will be applied� These values can be set separately�
The "CW time" and "CCW time" can be set between 00:00 [mm:ss] to 10:00  [mm:ss]�
The "Stop time" can be set between 00:05 [mm:ss] to 10:00  [mm:ss]�

Note: 

When either of "CW time" or "CCW time" is set to 00:00 [mm:ss], the other one can 

only set to minimum 00:03 [mm:ss]�

Symbol Rotation 

Direction

Graph

Description

 CW

0

CW

rpm

time

CCW

Factory setting: 

Continuous mode in clockwise�

 CW

0

CW

300

rpm

time

300

CCW

"Intermittent mode" is activated and 
"CW time" / "Stop time" is set�

Note:

 "CW time" and "Stop time" can be 

set separately�

  

 CCW

0

CW

300

rpm

time

300

CCW

"Intermittent mode" is activated and 
"CCW time" / "Stop time" is set�

Note:

 "CCW time" and "Stop time" can be 

set separately�

CCW / CW

0

CW

300

rpm

time

300

CCW

"Intermittent mode" is activated and 
"CW time" / "CCW time" / "Stop time" 
is set�

Note:

 "CW time" / "CCW time" and "Stop 

time" can be set separately�

/// Menu structure

Menu

Factory setting

C ...

A B

Mode

A

-

B

-

C

Safety

Password

000

-

Program

Program 1

-

Program 2

-

Program 5

-

...

Tempering

Control rate

Fast

-

Moderate

-

Slow

Mixing

Intermittent mode

-

Activate

00:10 [mm:ss]

CW time
Stop time

00:10 [mm:ss]

CCW time

00:00 [mm:ss]

Graph

Axis scaling

Automatic

00:00:00 [hh

:

mm:ss]

Timer

Mixing

Timer / Counter

00:00:00 [hh:mm:ss]

Tempering

Timer / Counter
Start Condition

Immediately

Set Temperature Reached

-

Mixing 

Axis assignment

Temperature

Minimum

Minimum

3000 rpm

100 ºC

Manual

Y-Axis 1: Speed

Y-Axis 2: Temperature

Maximum

Maximum

0 rpm

1.0 ºC

White

-

Firmware Update Information

°F

-

Display

Background

Black

Sound

Key Tone

Factory Settings

Firmware Update Information

yes

Mixing

00:10 [mm:ss]

Languages

Settings

English
Deutsch 

-

...

-

Units

°C

Temperature

Control rate

Tempering

Fast

Version

Information

Software

1.02.003/1.01.021

CW time

00:00 [mm:ss]

CCW time
Stop time

00:10 [mm:ss]

Summary of Contents for MATRIX ORBITAL DELTA PLUS

Page 1: ...IKA MATRIX ORBITAL DELTA PLUS...

Page 2: ...2 3 ENGLISH...

Page 3: ...rning symbols 6 Safety instructions 7 Unpacking 10 Operation 17 Intended use 9 Technical Data 38 Interfaces and outputs 31 Warranty 39 Maintenance and cleaning 37 Accessories 37 Operator panel and dis...

Page 4: ...arting up and follow the safety instructions Keep the operating instructions in a place where it can be accessed by everyone Ensure that only trained staff work with the device Follow the safety instr...

Page 5: ...foreign objects liquids and other contaminants Never use tubes made of glass or other fragile material Glass tubes can be smash Only use sealed micro test tube and sealed microplate deepwell plate Th...

Page 6: ...empering function E Timer button Enter the timer menu F Programs Prog button Enter the programs menu G Menu button Touch it once main menu is displayed Touch it a second time back to the working scree...

Page 7: ...temperature reached see Tempering under Timer Timer Counter symbol Mixing Indicate the Timer Counter for mixing Mixing symbol Indicate the mixing function is activated 00 10 Mixing in Intermittent mod...

Page 8: ...14 15 Inserting the deepwell plate DWP microplate MTP Inserting the micro test tubes Attaching the exchangeable attachments 1 2 1 2 1 Click 3 2 A B C D E F G H 1 2...

Page 9: ...een established with the device by pressing down the button the heating function can prevent the formation of condensate Operation Switching on IKA MATRIX ORBITAL Delta Version 1 02 003 1 01 021 Shake...

Page 10: ...tual rpm 25 1 0 300 0 00 00 0 00 01 38 0 LIMIT 1 0 LIMIT 100 0 0 00 00 99 59 59 25 0 38 0 max Setting the speed and starting mixing A IKA MATRIX 2 0 mL Fast actual C Set C Set rpm actual rpm 38 0 38 0...

Page 11: ...s 3 3 3 3 3 3 F F F F s s s s s s s s 2 2 2 Restarting the tempering and mixing function Note During the pause status the tempering mixing function can be switched off separately by touching temperin...

Page 12: ...0 mL Fast actual C Set C Set rpm actual rpm 38 0 38 0 500 500 0 00 39 0 00 50 A IKA MATRIX 2 0 mL Fast actual C Set C Set rpm actual rpm 38 0 500 0 00 40 0 00 51 38 0 500 Unlocking the buttons 1s A I...

Page 13: ...0 Fast 0 00 00 10 00 00 0 500 Quick mix Note When the tempering function is activated the Quick mix function is not available Press the Menu button G Select right menu item by touching the speed butt...

Page 14: ...ed and CW time Stop time is set Note CW time and Stop time can be set separately CCW 0 CW 300 rpm time 300 CCW Intermittent mode is activated and CCW time Stop time is set Note CCW time and Stop time...

Page 15: ...atically executed In control mode x x K The hysteresis tolerance of the actual temperature to target temperature is set e g 0 1 K The target values and settings of the segment are valid until the actu...

Page 16: ...t C or F for the temperature A check mark indicates the unit that is set for the system Display In the Background menu option you select black or white color for the working screen In the Firmware Upd...

Page 17: ...ed to the WD safety temperature limit value The error message Error 2 is displayed The above details correspond as far as possible to the recommendations of the NAMUR working party NAMUR recommendatio...

Page 18: ...ons between device and external devices PC 1 1 Cable This cable is required to connect RS 232 port to a PC 1 2 RxD 3 TxD 4 5 GND 6 7 RTS 8 CTS 9 1 RxD 2 TxD 3 4 GND 5 6 RTS 7 CTS 8 9 PC 1 2 3 4 5 6 7...

Page 19: ...r error code is displayed then take one of the following steps Contact the service department Send the device for repair including a short description of the fault Error 49 Motor hall sensor fault Err...

Page 20: ...e delivery invoice and giving reasons for the claim You will be liable for freight costs The warranty does not cover worn out parts nor does it apply to faults resulting from improper use insufficient...

Page 21: ...sales england ika com MALAYSIA IKA Works Asia Sdn Bhd Phone 60 3 6099 5666 eMail sales lab ika my CHINA IKA Works Guangzhou Phone 86 20 8222 6771 eMail info ika cn POLAND IKA Poland Sp z o o Phone 48...

Reviews: