21
Please unpack the equipment carefully and check for any damages. It
is important that any damages which may have arisen during transport
are ascertained when unpacking. If applicable a fact report must be
sent immediately (post, rail or forwarder).
The delivery scope covers:
A KS 130 basic or KS 130 control, for attachment screws, a single-head
spanner, a connection cable and operating instructions.
With the purchase of this device, you have acquired a high-quality
product. The design of the unit and its special shape ensure ease of
handing and problem-free work. The materials used and an exact
designation of them will significantly facilitate and simplify recycling
and reuse of parts.
The speed-controlled external rotor asynchronous motor allows for
infinitie speed adjustment in the range from 0 to 800 rpm. Electronic
motor control holds the set speed constant even if the weight of the
material on the surface increases. The heat given off by the motor can
be used to heat up the holding surface for the agitation containers.
The Control variant of the device is equipped with a locking device. This
makes it possible to fasten the agitation table in a defined position. By
removing the rubber feet, the device can be connected or screwed in
place in a defined position or with a system located next to it. Make
certain that the maximum depth of 5 mm for inserting the screws is
not exceeded.
The Control variant of the device can be delivered with a reverse opera-
ting mode (clockwise / counterclockwise motion) for the vibration table
on request. However, the reverse operation mode can only be used by
means of the serial interface in combination with a PC (for example
with labworld
soft
®
). If the device is not operated on a serial interface,
it behaves like a Control device without reverse operation mode.
It is easy to exchange the various attachments.
The Control device variant is equipped with a serial interface that
makes it possible to control the device through the PC (for example
with labworld
soft
®
.
Check whether the voltage specified on the type plate matches the
mains voltage available. The power socket used must be earthed
(protective earth conductor contact). If these conditions are met, the
device is ready to operate after plugging in the mains plug. If these
procedures are not followed, safe operation cannot be guaranteed
and/or the equipment may be damaged.
Observe the ambient conditions (temperature, humidity, etc.) listed
under Technical Data.
When you turn on the basic device variant, the turn button for the time
setting must be in the middle setting (not activated).
If the mounting surface is not even, you can improve the safety level
of the device in reference to how it stands with the adjustable feet. To
do this, turn the appropriate device foot downward with the wrench
(included with delivery) until the device is standing securely on the
surface.
To prevent the foot of the
device from coming loose
by itself, hold down the
foot with one hand and
tighten the oppsite hexa-
gonal nut with the
other hand until it is finger
tight.
Unpacking
Useful facts
Commissioning
IKA
R
1/min
min
Time
basic
HS 250
OFF
ON
10
5
15
Mot
60
30
90
25
20
30
35
45
40
50
210
120
150
180
270
240
300
A
Summary of Contents for KS 130 basic
Page 51: ...51...