Setting the safety temperature

Setting the safety temperature with screwdriver delivered 
with the device� 

Factory setting: approximate max� value
Adjustment range: 0 to 110 °C�

Warning: 

The safety temperature must always be set to at 

least 25 ºC lower than the flash point of the fluid used� 

Fig. 13

+

-

The safety temperature setting will appear on the display�

Safety temperature

Note: 

Before commissioning, make sure that the device has 

not been moved in one hour�

Check whether the voltage specified on the type plate 
matches the power voltage available�

The power socket used must be 
earthed (protective earthing)�

If above conditions are met, the instrument is ready for op-
eration after plugging in the power plug�
If these conditions are not met, safe operation is not guar-
anteed and the device could be damaged�

Observe  the ambient conditions (temperature, humidity, 
etc�) listed under “Technical Data”�

After switching on the power switch (10) at the back of the 
device or pressing the on/off button (A), all LED segments 
light up during the self test� 

Commissioning and operating

After that, following two screens appear alternately�

Display field

Software version (ver)

Operating mode (mode)

Working setting

Then the device enters standby status and is ready for op-
eration�

Change the temperature setting with “Temp (+)“ button (M) 
or “Temp (-)“ button (N)�
Change the pump speed setting with “Pump (+)“ button (J) 
or “Pump (-)“ button (K)�

In standby status, activate the tempering function by press-
ing the “Temp“ button (L)� The device starts tempering 
function and pumping function� 
After pressing the power switch (10) the start-up of the 
compressor is delayed (about 3 minutes)�

In working status, press the “OK/Pump“ button (I) to stop 
pumping and tempering function�

Note:

 In running status, the pump speed adjusts automati-

cally according to the load� When the actual speed doesn’t 
reach the setting speed, the blinking set speed and the ac-
tual speed appears alternately�

In standby status, press the “OK/Pump“ button (I) to start 
the pump function� The tempering function will not be ac-
tivated�
In working status, press the “Temp“ button (L) to stop the re 
tempering function, the pump keep running�

Pressing the lock button (H) for about 2 seconds, the device 
control elements can be locked to prevent accidental chang-
es that can be done during operation (LED (G) will light up)�
To unlock the control elements, press the lock button again 
for about 2 seconds (LED (G) goes out)� 

Safety temperature

Summary of Contents for HRC 2 basic

Page 1: ...20000015724a IKA HRC 2 basic HRC 2 basic_112017 Operating instructions EN...

Page 2: ...l and display 2 Handle 3 Venting grid 4 Filling opening lid 5 RS 232 port 6 USB port 7 Overflow 8 Backflow 9 Power socket 10 Power switch 11 Pump connector OUT 12 Pump connector IN Fig 1 1 3 4 2 2 2 2...

Page 3: ...ard information for IKA fluid 9 Moving the device 10 Operator panel and display 10 Setting the safety temperature 11 Commissioning 11 Menu settings 12 Menu structure 12 Basic guide to setting the menu...

Page 4: ...Explication of warning symbols Indicates an extremly hazardous situation which if not avoided will result in death serious injury Indicates a hazardous situation which if not avoided can result in de...

Page 5: ...t inhibit fire or explo sions or comprehensive monitoring equipment Process pathogenic material only in closed vessels under a suitable fume hood Please contact IKA application sup port if you have an...

Page 6: ...or chromium salts Glycerine Ferrous water When changing the bath fluid type must remove the remaining water from the complete system includ ing hoses and external devices When doing this also open th...

Page 7: ...of heat sources and do not expose to sun light Cooling machine pump motor and electronics produce intrinsic heat that is dissipated via the venting grids 3 Never cover these venting grids Note After s...

Page 8: ...e to the drain port and turn the drain valve in counter clockwise direction with a straight screwdriver Red Low Level Red High Level Low Level High Level Fig 9 Warning Do not empty the bath while the...

Page 9: ...distilled or bi distilled water are unsuitable for operation due to corrosive properties of these media High purity water and distillates are suitable as a medium after adding 0 1 g soda Na2CO3 sodium...

Page 10: ...rigerating The LED indicates the refrigerating function is activated E LED display Display the settings and actual values F Menu button Press it once menu option is display Press it once again back to...

Page 11: ...rnately Display field Software version ver Operating mode mode Working setting Then the device enters standby status and is ready for op eration Change the temperature setting with Temp button M or Te...

Page 12: ...uide to setting the menu options F Enter the menu by pressing Menu button F F Press the Temp button M or Temp button N to change the menu options F Press the Pump button J or Pump button K to change t...

Page 13: ...riable Ti values that are too small can lead to instability of the controller The differential time Td s is the derivative time and de termines how strongly the rate of change of the control deviation...

Page 14: ...ing operation hot plugging Con nected devices and their properties are automatically recog nized The USB port can also be used to update firmware USB device driver First download the latest driver for...

Page 15: ...soft provides a convenient tool for controlling circulating equipment and collecting data under MS Windows and includes graphical entry features for pump motor speed ramps for example Commands NAMUR C...

Page 16: ...ce Dirt Cleaning agent Dye Isopropyl alcohol Construction material Water containing tenside isopropyl alcohol Cosmetics Water containing tenside isopropyl alcohol isopropyl alcohol Water containing te...

Page 17: ...m Check the bath temperature measurement Err 13 Pump off Heating refrigerating off Heater switched off by safety circuit Check safety temperature set value fluid level Err 15 Pump off Heating refriger...

Page 18: ...eating capacity W 1500 230 VAV 1200 115 VAC 910 100 VAC Cooling capacity according to DIN 12876 at 4000 rpm 20 C 10 C 0 C 10 C 20 C W 400 370 320 240 130 Refrigerant R134a Refrigerant quantity g 230 M...

Page 19: ...ill be liable for freight costs The warranty does not cover worn out parts nor does it ap ply to faults resulting from improper use insufficient care or maintenance not carried out in accordance with...