12
Menu settings
•
Maximum temperature (HI T):
The maximum adjustable value: 100 °C�
This value can be limited additionally by the selected fluid
(No�, see “Fluid (Standard information for
IKA
®
fluid)”�
Note:
The maximum value is limited by the set safety
temperature�
•
Minimum temperature (LO T):
The minimum adjustable value: - 20 °C
�
•
Maximum speed (HI R):
The maximum adjustable value:
4000 rpm
�
•
Minimum speed (LO R):
The minimum adjustable value: 2000 rpm�
•
Operating mode (MODE):
Mode A:
After power on/power failure no automatic re-
start of functions�
Mode B:
After power on/power failure automatic restart
of functions, depending on previous settings�
Mode C:
Set values (set in mode A or B) cannot be
changed� After power on/power failure automatic restart
of functions, depending on previous settings�
•
Basic guide to setting the menu options:
F
Enter the menu by pressing ”Menu” button (F)�
F
Press the“Temp (+)“ button (M) or “Temp (-)“ button (N)
to change the menu options�
F
Press the“Pump (+)“ button (J) or “Pump (-)“ button (K)
to change the menu value settings�
F
Confirm the menu settings by pressing the “OK/Pump“
key (I)�
MODE
Maximum temperature (HI T)
Minimum temperature (LO T)
Maximum speed (HI R)
Minimum speed (LO R)
Fluid type (FLUI)
The type of temperature PID control arithmetic (AUTO)
(Heating)
Alarm and button tone (BEEP)
Calibration (CALI)
Menu
• Menu structure:
Default settings
A
B
C
Proportional coefficient of PID (Kp 1)
Integrate time of PID (Ti 1)
Differential time of PID (Td 1)
AUTO 1
AUTO 0
BEEP 0
BEEP 1
CALI 0
CALI 2
CALI 3
Activated
-
-
100 °C
- 20 °C
4000 rpm
2000 rpm
8
Activated
60�0
5�00
0�3
-
Activated
Activated
-
-
•
Fluid type
(
FLUI
):
The chosen fluid (No�) limits the maximum and minimum
temperature� Customized fluid (No�: 8) enables the whole
working temperature range� See Section “Fluids (Stan-
dard information for
IKA
®
fluid)”�