background image

14

Set the first point temperature (step 1, e�g� 10 °C) with 
“Pump (+)” button (J) or “Pump (-)” button (K)� Confirm 
the setting by pressing “OK/Pump“ button (I)� 

Left area of the display appears the set value (10 °C) and 
the right area of the display appears the actual tempera-
ture value measured� The “Set” LED (C) flashes briefly� 

The unit now starts and controls to the set value� When 
the temperature has reached the set value and has be-
come constant, the “Set” LED (C) no longer flashes and 
the following display appears� The temperature measured 
by the unit appears on the left hand side�

Input the calibration value from the reference measuring 
instrument (e�g� 10�1 °C) with “Pump (+)” button (J) or 
“Pump (-)” button (K)�

Confirm the value by pressing the “OK/Pump“ button (I)�
Back to previous screen for input a new value by pressing 
“Temp” button (L)�

Back

OK

The first point calibration is finished now�
Calibration of the other points is performed in the same 
way�

Example: 2-point calibration:

Dip the temperature sensor of the reference measuring 
instrument into the bath fluid�

Select 2-point calibration in the menu� Press the “OK/
Pump“ button (I) to start the 2-point calibration�

 Interface and output

The device can be operated in “Remote” mode via the RS 232 
port (5) or the USB port (6) connected to a PC and with the 
laboratory software Labworld

soft

®

Note: 

Please observe the system requirements together 

with the operating instructions and help section included 
with the software�

USB interface:

The Universal Serial Bus (USB) is a serial bus for connecting 
the device to the PC� Equipped with USB devices can be 
connected to a PC during operation (hot plugging)� Con-
nected devices and their properties are automatically recog-
nized� The USB port can also be used to update firmware�

USB device driver:

First, download the latest driver for 

IKA

®

 devices with USB 

port from: 

http://www.ika.com/ika/lws/download/usb-driver.zip.

 

Install the driver by running the setup file� Then connect the 

IKA

®

 device through the USB data cable to the PC� 

The data communication is via a virtual COM port� Configu-
ration, command syntax and commands of the virtual COM 
ports are as described in RS 232 interface�

Device software update:

For device software update, visit 

IKA

®

 website 

www.ika.

com

 and enter the “Service” menu� Download and run the 

Firmware Update Tool�

Find and click the Firmware Update Tool in your PC after 
installation� Register your E-mail and password� 
Connect the device to your PC via the USB cable� Then, you 
can update the device software according to instructions of 
the Firmware Update Tool�

RS 232 interface:

Configuration:
- The functions of the interface connections between the 

stirrer machine and the automation system are chosen 
from the signals specified in EIA standard RS 232 in ac-
cordance with DIN 66 020 Part 1�

- For the electrical characteristics of the interface and the al-

location of signal status, standard RS 232 applies in accor-
dance with DIN 66 259 Part 1�

- Transmission procedure: asynchronous character transmis-

sion in start-stop mode�

- Type of transmission: full duplex�
-  Character format: character representation in accordance 

with data format in DIN 66 022 for start-stop mode� 1 
start bit; 7 character bits; 1 parity bit (even); 1 stop bit�

- Transmission speed: 9600 bit/s�
- Data flow control: none
- Access procedure: data transfer from the stirrer machine to 

the computer takes place only at the computer’s request�

Summary of Contents for HRC 2 basic

Page 1: ...20000015724a IKA HRC 2 basic HRC 2 basic_112017 Operating instructions EN...

Page 2: ...l and display 2 Handle 3 Venting grid 4 Filling opening lid 5 RS 232 port 6 USB port 7 Overflow 8 Backflow 9 Power socket 10 Power switch 11 Pump connector OUT 12 Pump connector IN Fig 1 1 3 4 2 2 2 2...

Page 3: ...ard information for IKA fluid 9 Moving the device 10 Operator panel and display 10 Setting the safety temperature 11 Commissioning 11 Menu settings 12 Menu structure 12 Basic guide to setting the menu...

Page 4: ...Explication of warning symbols Indicates an extremly hazardous situation which if not avoided will result in death serious injury Indicates a hazardous situation which if not avoided can result in de...

Page 5: ...t inhibit fire or explo sions or comprehensive monitoring equipment Process pathogenic material only in closed vessels under a suitable fume hood Please contact IKA application sup port if you have an...

Page 6: ...or chromium salts Glycerine Ferrous water When changing the bath fluid type must remove the remaining water from the complete system includ ing hoses and external devices When doing this also open th...

Page 7: ...of heat sources and do not expose to sun light Cooling machine pump motor and electronics produce intrinsic heat that is dissipated via the venting grids 3 Never cover these venting grids Note After s...

Page 8: ...e to the drain port and turn the drain valve in counter clockwise direction with a straight screwdriver Red Low Level Red High Level Low Level High Level Fig 9 Warning Do not empty the bath while the...

Page 9: ...distilled or bi distilled water are unsuitable for operation due to corrosive properties of these media High purity water and distillates are suitable as a medium after adding 0 1 g soda Na2CO3 sodium...

Page 10: ...rigerating The LED indicates the refrigerating function is activated E LED display Display the settings and actual values F Menu button Press it once menu option is display Press it once again back to...

Page 11: ...rnately Display field Software version ver Operating mode mode Working setting Then the device enters standby status and is ready for op eration Change the temperature setting with Temp button M or Te...

Page 12: ...uide to setting the menu options F Enter the menu by pressing Menu button F F Press the Temp button M or Temp button N to change the menu options F Press the Pump button J or Pump button K to change t...

Page 13: ...riable Ti values that are too small can lead to instability of the controller The differential time Td s is the derivative time and de termines how strongly the rate of change of the control deviation...

Page 14: ...ing operation hot plugging Con nected devices and their properties are automatically recog nized The USB port can also be used to update firmware USB device driver First download the latest driver for...

Page 15: ...soft provides a convenient tool for controlling circulating equipment and collecting data under MS Windows and includes graphical entry features for pump motor speed ramps for example Commands NAMUR C...

Page 16: ...ce Dirt Cleaning agent Dye Isopropyl alcohol Construction material Water containing tenside isopropyl alcohol Cosmetics Water containing tenside isopropyl alcohol isopropyl alcohol Water containing te...

Page 17: ...m Check the bath temperature measurement Err 13 Pump off Heating refrigerating off Heater switched off by safety circuit Check safety temperature set value fluid level Err 15 Pump off Heating refriger...