20
You have purchased an original IKA laboratory machine which
meets the highest engineering and quality standards.
In accordance with IKA guarantee conditions, the guarantee peri-
od is 24 months. For claims under the guarantee please contact
your local dealer. You may also send the machine direct to our
works, enclosing the delivery invoice and giving reasons for the
claim. You will be liable for freight costs.
Construction in accordance with the foolwing safety standards
EN 61 010-1 UL 3101-1
CAN/CSA C22.2 (1010-1)
Construction in accordance with the foolwing EMC standards
EN 50 082-1 EN 55 014 EN 60 555-2; -3 EN 50 081
Associated EU guidlines
EMC-guidlines: 89/336/EWG
Machine guidelines: 73/023/EWG
Design voltage:
VAC
230±
10%
oder
VAC
100 115±
10%
Design frequency:
Hz
50/60
Power consumption:
W
300
Output power:
W
160
Motor speed, idle:
1/min
28 000
Motor speed, under load:
1/min
19 500
A 11 022000
Guarantee
Associated standards and regulations
Technical data
Circumferal speed:
Idle
m/s
76
Under load
m/s
53
Effective volume:
A11.5 Grinding container
ml
80
A11.4 Grinding container
ml
250
Working time:
Short-time operation (KB)
min
5 / 10
5 min. ON / 10 min OFF
Safety:
Safety switch on the grinding
container, temperature limi-
ter in the motor
Fuse on power board
230V:
A
8
Id.-Nr. 31 406 00
115V:
A
10
Id.-Nr. 31 407 00
Malfunction display:
Orange-red glow lights,
display of over-temperature
and grinding area not closed
Drive:
Universal motor
Permitted ambient
temperature:
°C
+5 to +40
Permitted relative humidity:
%
80
Protection type to DIN 40 050:
IP 43
Protection class:
I
Excess voltage category:
II
Contamination level:
2
Noise level:
dB
max. 85
Operation at a terrestrial
altitude
m
max. 2000 above sea level
Dimensions: (W x D x H)
mm
85 x 85 x 240
Weight:
kg
1,5
A 11 23.04.2001 8:10 Uhr Seite 20