
6
deviation. Please consult your physician about the result.
* Don’t kink the connection tube during use, otherwise, the cuff pressure may continu
-
ously increase which can prevent blood flow and result in harmful injury to the PATIENT.
* When using this device, please pay attention to the following situation which may inter
-
rupt blood flow and influence blood circulation of the patient, thus cause harmful injury
to the patient: connection tubing kinking too frequent and consecutive multiple measure-
ments; the application of the cuff and its pressurization on any arm where intravascular
access or therapy, or an arterio-venous (A-V) shunt, is present; inflating the cuff on the
side of a mastectomy.
* Warning: Do not apply the cuff over a wound;otherwise it can cause further injury.
*Do not inflate the cuff on the same limb which other monitoring ME equipment is applied
around simultaneously, because this could cause temporary loss of function of those
simultaneously-used monitoring ME equipment.
*On the rare occasion of a fault causing the cuff to remain fully inflated during measure
-
ment, open the cuff immediately. Prolonged high pressure (cuff pressure
>
300mmHg
or constant pressure > 15mmHg for more than 3 minutes) applied to the arm may lead
to an ecchymosis.
*Please check that operation of the device does not result in prolonged impairment of
patient blood circulation.
* When measurement, please avoid compression or restriction of the connection tubing.
* The device cannot be used with HF surgical equipment at the same time.
* The ACCOMPANYING DOCUMENT shall disclose that the SPHYGMOMANOMETER
was clinically investigated according to the requirements of ISO 81060-2:2013.
* To verify the calibration of the AUTOMATED SPHYGMOMANOMETER, please contact
the manufacturer.
* This device is contraindicated for any female who may be suspected of, or is pregnant.
Besides providing inaccurate readings, the effects of this device on the fetus are unknown.
* Too frequent and consecutive measurements could cause disturbances in blood
circulation and injuries.
* This unit is not suitable for continuous monitoring during medical emergencies or op
-
erations. Otherwise, the patient’s arm and fingers will become anaesthetic, swollen and
even purple due to a lack of blood.
* When not in use, store the device in a dry room and protect it against extreme moisture,
heat, lint, dust and direct sunlight. Never place any heavy objects on the storage case.
* This device may be used only for the purpose described in this booklet. The manufac
-
turer cannot be held liable for damage caused by incorrect application.
*This device comprises sensitive components and must be treated with caution.
Observe the storage and operating conditions described in this booklet.
* The equipment is not AP/APG equipment and not suitable for use in the presence of a
flammable anesthetic mixture with air of with oxygen or nitrous oxide.
* Warning: No servicing/maintenance while the ME equipment is in use.
* The patient is an intended operator.
* The patient can measure data and change batteries under normal circumstances and
maintain the device and its accessories according to the user manual.
* To avoid measurement errors, please avoid the condition of strong electromagnetic
field radiated interference signal or electrical fast transient/burst signal.
* The blood pressure monitor and the cuff are suitable for use within the patient environ
-
ment. If you are allergic to polyester, nylon or plastic, please don’t use this device.
* During use, the patient will be in contact with the cuff. The materials of the cuff have