NAVAIR 17—2OAX—737
2.2 Electronic
counter
2.3 Digital multimeter
(DMM)
2.4 Power
meter/power
sensor
2.5 Signal generator
calibrator
2.7 Navigation
indicator T/S
2.8 Spectrum
analyzer
2.9 Connector
gage kit
*Non_NCE equipment
Bandwidth: 50 MHz
Uncertainty: ±3%time base
Dual trace and delayed
sweep capability
Frequency range: 25 Hz to 335 MHz
Uncertainty: ±0.25ppm
Resolution: 1 Hz
DC voltage range: 0 to ±80mV dc
Uncertainty: ±1%
Power range: —25 to —35 dBm
Uncertainty: ±1.25%
Frequency range: 70 to 335 MHz
Frequency range: 108 to 118 MHz
Tuned RF level rang: e—6 to —100 dBm
Uncertainty: ±0.35dB
AM modulation range: 15% to
100% AM
Uncertainty: ±1%AM
FM modulation range: <200 Hz to
400 Hz FM
Uncertainty: ±2%
Capable of measuring distortion down to
0.1% with 0.01% resolution
No known substitute
Range: 50 MHz
to 1.2 GHz
Resolution: 300 Hz mm
Dynamic range:
>85
dB
Range: 0 to 0.197 inch
Uncertainty: ±0.0025inch
(MIL—C--71B)
Tektronix
7704AM0D 1 29GOPTO3,
7904AOPT03 or
7704AM0D129G with
7A26 and 7B92A plug—ins
Hewlett—Packard
5335A0PT010, 030, 040,
5345A0PT012, or 5345A
Fluke 8SO6AAN, 8502AAT,
or 884OAAFOPTO5
Hewlett—Packard
436A0PT022, 436A, or
435A with 8484A
Hewlett—Packard
8902A0PTE02 (consists
of: H-P 8902A0PT002;
and H—P 11722A) or
8902A0PTE04 (consists
of: H—P8902A;
H—P 8672AOPT001, 008;
H—P 11792A0PT001,H04;
and H—P 1 1793A0PTH04)
Sound Technology
1700BOPTOO3, 005; or
Hewlett—Packard
8903B0PT050, 907, 910,
or 8903BOPT907
Collins 478A—3 ZIFOR 111*
Hewlett—Packard
8562A0PTE50,
8569BOPT001, or 8566B
Maury Microwave
AOO7B, AOO7A
2.1 Oscilloscope
Item
Minimum Use
Specifications
Calibration
Equipment
2.6 Distortion
analyzer
5