NAVAIR 17—2OAX—737
4.2.12 Set the TI FREQUENCY MHz switches to the following settings. At each setting, verify that the
power meter indicates within
the tolerance limits listed.
TI
FREQUENCY MHz
Switch Setting
Power Meter
Tolerance Limits
(dBm)
072.000
074.000
076.000
079.000
—28.5
to —31.5
“
“
“
4.2.13 Disconnect the power sensor from the TI.
4.3 ATTENUATOR LINEARITY TESTS
4.3.1 Set the TI controls as follows:
FREQUENCY MHz switches to
MASTER MOD control to
108.000 MHz
UNCAL [fullyccw (0)]
Output attenuator control to
-6 dBm
4.3.2 Connect the signal generator calibrator sensor module input to the TI RF output connector.
4.3.3 Press the signal generator calibrator INSTR PRESET key (Blue key, AUTOMATIC OPERATION
key), and ensure that the signal generator calibrator is in the FREQ measurement mode and tunes to the TI
FREQUENCY MHz switches setting. Set the signal generator calibrator for TUNED RF LEVEL
measurements with log (dBm) display, enter 39.9 SPCL, and then press the CALIBRATE key.
4.3.4 Set the TI output attenuator control to each of the following settings. At each setting, verify that the
signal generator calibrator indicates within the tolerance limits listed.
TI Output Attenuator
Control Setting
Signal Generator Calibrator
Tolerance Limits
(dBm)
—6
—4.5
to
—7.5
—10
—8.5
to
—11.5
—20
—18.5
to
—21.5
—30
—28.5
to
—31.5
—40
—38.5
to
—41.5
—50
—48.5
to
—51.5
—60
—57.5
to
—62.5
—70
—67.5
to
—72.5
—80
—77.5
to
—82.5
—90
—87.5
to
—92.5
—100
—97.5
to —102.5
4.3.5 Disconnect the signal generator calibrator sensor module from the TI.
9