10.2 Self-diagnostics / fault indications
The unit has many self-diagnostic options�
• It monitors itself automatically during operation�
• Warnings and faults are displayed (even if the display is deactivated), in
addition they are available via IO-Link�
Display
Status LED
OUT1
Status LED
OUT2
Type of fault *
)
Fault / warning
Corrective measures
PARA
F Parameter setting outside the
permissible range�
►
Repeat parameter setting�
none
F Supply voltage too low�
►
Check / correct the supply
voltage�
SC
Flashes
Flashes Flashes F Excessive current on
switching outputs OUT1 and
OUT2 **
)
�
►
Check switching
outputs for short circuit
or excessive current;
Remove the fault�
SC1
Flashes
Flashes
F Excessive current at
switching output OUT1 **
)
�
►
Check switching output
OUT1 for short-circuit
or excessive current;
Remove the fault�
SC2
Flashes
Flashes F Excessive current at
switching output OUT2 **
)
�
►
Check switching output
OUT2 for short-circuit
or excessive current�
Remove the fault�
Loc
W Parameter setting locked via
buttons�
►
unlocking"�
C�Loc
W Parameter setting locked
via pushbuttons, parameter
setting is enabled via IO-Link
►
Wait until parameter
setting via IO-Link is
finished�
S�Loc
W Setting buttons locked
via parameter software�
Parameter change is rejected
►
Unlocking only possible
via IO-Link interface
/ parameter setting
software�