23
UK
►
Select [cOP]
►
Press [SET] and keep it pressed.
> [cOP] flashes for some seconds; then the continuous display indicates
that the adjustment is being made.
> If the adjustment is successful, [donE] is displayed.
►
Confirm with [Mode/Enter].
> If the adjustment is not successful, [FAIL] is displayed.
►
Lower the level if necessary or correct the position of the overflow
protection [OP] and repeat the adjustment operation.
cOP
Minimum distance between the overflow protection OP and the level during
adjustment:
Table 10-1
[cm]
[inch]
LKx022
2.0
0.8
LKx023
3.5
1.4
LKx024
5.0
2.0
The position of the overflow protection OP can be determined by calling up
the parameter [OP]. Note the offset if necessary.
The current level is to be determined manually since the unit is not yet
ready for operation before the adjustment.
With the setting [MEdI] = [Auto] or [OP] = [OFF], the parameter [cOP] is not
available.
When the overflow protection is activated ([OP] = [value...]), an adjustment
[cOP] must be carried out each time:
• [MEdI] or [OP] has been changed. In this case
appears in the
display.
• the installation position (height, orientation) has been changed.
• the connection between the sensor and the tank ground (e.g. length of
the jumper cable) has been changed.
With deactivated overflow protection [OP] = [OFF] or [MEdI] = [Auto]:
To assign the basic settings and to adapt to the medium and installation
environment the unit has to be reinitialised when installed.
►
Switch the operating voltage off and on again.