background image

5

HRRS Series

5

Operation

Chapter 3  

OPERATION

and a setting of 10-10-10-l0-10.10 

Ω

 becomes:

 

   

1 0  

1  0   0   0   0   0 . 0

 

    1 0  

  1   0   0   0   0 . 0

 

    1 0  

    1   0   0   0 . 0

 

    1 0  

      1   0   0 . 0

 

    1 0  

        1   0 . 0

 

  . 1 0  

          1 . 0        

      TOT 

1 1 1 1 1 1.0

3.3  Environmental Conditions

For optimal accuracy, the decade box should be used 

in an environment of 23°C. It should be allowed to 

stabilize  at  that  temperature  after  any  significant 

temperature variation.

Humidity should be maintained at <50% RH. This is 

especially important if high resistances are involved.

3.4  Storage

If this instrument is to be stored for any lengthy period 

of time, it should be sealed in plastic and stored in a 

dry location.  It should not be subjected to temperature 

extremes beyond the specifications.  Extended expo-

sure to such temperatures can result in an irreversible 

change in resistance, and require recalibration.

3.1  Initial Inspection and Setup

This instrument was carefully inspected before ship-

ment. It should be in proper electrical and mechanical 

order upon receipt.

An 

OPERATING GUIDE

 is attached to the case 

of  the  instrument  to  provide  ready  reference  to 

specifications.

3.2  Dial Setting

Whenever  the  dials  are  used in positions  0-9, the 

resulting resistance is read directly. Both the decimal 

point and the steps are clearly marked on the panel.

For  additional  flexibility  and  range,  each  decade 

provides a “10” position setting. This “10” position 

on any one decade equals the “1” position on the next 

higher decade. It adds about 11% to the nominal total 

decade resistance.

To determine the resistance obtained when one or 

more “10” settings are used, simply add “1” to the 

next higher decade. For example, a setting of  3-6-

10-0-10 Ω becomes:

 

   

3  

3  0   0   0   0

 

    6  

  6   0   0   0

 

  1 0  

  1   0   0   0

 

    0  

      0   0

 

  1 0  

      1   0        

 

 TOT 

3 7 0 1 0

Summary of Contents for HRRS SERIES

Page 1: ...l info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC HRRS SERIES High Resistance Decade Substituter User and Service Manual Copyright 2013 IET Labs Inc Visit www ietlabs c...

Page 2: ...Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC...

Page 3: ...placed at no charge when returned to IET Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the specifications shall void this warranty IE...

Page 4: ...45 V ARE USED TAKE ALL MEASURES TO AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS WHEN USING THIS INSTRUMENT Use extreme caution when...

Page 5: ...Storage 5 Chapter 4 Maintenance 4 1 Maintainability and Reliability 6 4 2 Preventive Maintenance 6 4 3 Verification of Performance 6 4 3 1 Calibration Interval 6 4 3 2 General Considerations 6 4 3 3 C...

Page 6: ...iv HRRS Series Table of Contents This page is intentionally left blank...

Page 7: ...applied input for each decade Higher voltage models up to 10 kV are also available from IET Labs The binding posts are insulated with Kel F high resistance non moisture absorbing material ensure low...

Page 8: ...0 1 1000 2000 15 1 100 1 G 10 G 0 20 0 50 1 1000 2000 50 1 500 High voltage film 10 G 100 G 0 50 1 1 1000 2000 50 2 500 100 G 1 T See HRRS 5kV Series if these values are required 1 T 10 T To apply up...

Page 9: ...racy replace with Q B or F as required Options RM Rack mountable case for standard 19 rack K Kelvin type 4 terminal posts RO Rear output 2 2 Ordering Information Model Total resistance Number of decad...

Page 10: ...nce Accuracy 18 28 C 50 RH Max voltage per step V Max Voltage V TC ppm C VC ppm V Stability ppm yr Resistor Type Q B F 10 100 0 01 2 m 0 03 2 m 0 10 2 m 2 5 25 15 10 Wirewound non inductive 100 1 k 0...

Page 11: ...sult in an irreversible change in resistance and require recalibration 3 1 Initial Inspection and Setup This instrument was carefully inspected before ship ment It should be in proper electrical and m...

Page 12: ...rtified calibration laboratory If the user should choose to perform this procedure then the considerations below should be observed 4 3 2 General Considerations It is important whenever testing the HR...

Page 13: ...ecifications on Page 2 4 Confirm that the resistances fall within these limits after subtraction of the zero resistance If any resistances fall outside thee limits the associated switch assembly may r...

Page 14: ...n HRRS SW 1k 1 k step Decade Switch Assembly Not Shown HRRS SW 10k 10 k step Decade Switch Assembly Not Shown HRRS SW 100k 100 k step Decade Switch Assembly Not Shown HRRS SW 1M 1 M step Decade Switch...

Page 15: ......

Reviews: