background image

6

HRRS Series

6

Maintenance

Chapter 4 

MAINTENANCE

4.1  Maintainability and Reliability

It is possible to maintain Model HRRS indefinitely.  It 

is reliable due to its closed design and sealed switches 

and resistors.  The unit is resistant to electromagnetic 

interference (EMI) because of its metal enclosure.

4.2  Preventive Maintenance

Keep the unit in a clean environment.  This will help 

prevent possible contamination.

The front panel should be periodically cleaned to 

eliminate any leakage paths from near or around the 

binding posts.  To clean the front panel:

Wipe the front panel clean using alcohol and a 

lint-free cloth

4.3  Verification of Performance

4.3.1  Calibration Interval

The 

HRRS 

Series instruments should be verified for 

performance at a calibration interval of twelve (12) 

months. This procedure may be carried out by the user 

if a calibration capability is available, by IET Labs, 

or by a certified calibration laboratory. 

If the user should choose to perform this procedure, 

then the considerations below should be observed.

4.3.2  General Considerations

It is important, whenever testing the HRRS Series 

Decade Units, to be very aware of the capabilities and 

limitations of the test instruments used. A resistance 

bridge may be employed, and there are direct-reading 

resistance  meters  or  digital  multimeters  available 

that can verify the accuracy of these units, especially 

when used in conjunction with standards that can 

serve to confirm or improve the accuracy of the test-

ing instrument

Such test instruments must have a 4:1 TUR  capabil-

ity for each value calibrated. Note that the accuracy 

varies for different decades. A number of commercial 

bridges and meters exist that can perform this task; 

consult IET Labs.

It is important to allow both the testing instrument 

and the HRRS Substituter to stabilize for a number of 

hours at the nominal operating temperature of 23°C, 

and at nominal laboratory conditions of humidity. 

There should be no temperature gradients across the 

unit under test.

Summary of Contents for HRRS SERIES

Page 1: ...l info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC HRRS SERIES High Resistance Decade Substituter User and Service Manual Copyright 2013 IET Labs Inc Visit www ietlabs c...

Page 2: ...Email info ietlabs com TEL 516 334 5959 FAX 516 334 5988 www ietlabs com IET LABS INC...

Page 3: ...placed at no charge when returned to IET Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the specifications shall void this warranty IE...

Page 4: ...45 V ARE USED TAKE ALL MEASURES TO AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE CONDUCTORS WHEN USING THIS INSTRUMENT Use extreme caution when...

Page 5: ...Storage 5 Chapter 4 Maintenance 4 1 Maintainability and Reliability 6 4 2 Preventive Maintenance 6 4 3 Verification of Performance 6 4 3 1 Calibration Interval 6 4 3 2 General Considerations 6 4 3 3 C...

Page 6: ...iv HRRS Series Table of Contents This page is intentionally left blank...

Page 7: ...applied input for each decade Higher voltage models up to 10 kV are also available from IET Labs The binding posts are insulated with Kel F high resistance non moisture absorbing material ensure low...

Page 8: ...0 1 1000 2000 15 1 100 1 G 10 G 0 20 0 50 1 1000 2000 50 1 500 High voltage film 10 G 100 G 0 50 1 1 1000 2000 50 2 500 100 G 1 T See HRRS 5kV Series if these values are required 1 T 10 T To apply up...

Page 9: ...racy replace with Q B or F as required Options RM Rack mountable case for standard 19 rack K Kelvin type 4 terminal posts RO Rear output 2 2 Ordering Information Model Total resistance Number of decad...

Page 10: ...nce Accuracy 18 28 C 50 RH Max voltage per step V Max Voltage V TC ppm C VC ppm V Stability ppm yr Resistor Type Q B F 10 100 0 01 2 m 0 03 2 m 0 10 2 m 2 5 25 15 10 Wirewound non inductive 100 1 k 0...

Page 11: ...sult in an irreversible change in resistance and require recalibration 3 1 Initial Inspection and Setup This instrument was carefully inspected before ship ment It should be in proper electrical and m...

Page 12: ...rtified calibration laboratory If the user should choose to perform this procedure then the considerations below should be observed 4 3 2 General Considerations It is important whenever testing the HR...

Page 13: ...ecifications on Page 2 4 Confirm that the resistances fall within these limits after subtraction of the zero resistance If any resistances fall outside thee limits the associated switch assembly may r...

Page 14: ...n HRRS SW 1k 1 k step Decade Switch Assembly Not Shown HRRS SW 10k 10 k step Decade Switch Assembly Not Shown HRRS SW 100k 100 k step Decade Switch Assembly Not Shown HRRS SW 1M 1 M step Decade Switch...

Page 15: ......

Reviews: