Error messages
37
EN
Error
description
Cause
Remedy
Err 4
Readings not transmitted
correctly.
Wait until you take another measurement. The data are then
transmitted together with the new readings.
If this error occurs repeatedly, send the blood pressure monitor to
your healthcare provider or directly to the manufacturer for
checking.
Err 5
Power pack or battery
voltage too low.
Change the batteries (
➔
Chapter 5.2
Battery contacts are
corroded.
Clean the battery contacts with a cotton cloth and a little alcohol.
Err 6
Air congestion
Check the blood pressure cuff for air congestion or a kink in the air
hose. If there is a kink in the air hose, unkink it, otherwise send the
blood pressure monitor to your healthcare provider or the
manufacturer.
Blood pressure cuff
incorrectly connected.
Connect the blood pressure cuff to the blood pressure monitor
(
➔
Chapter 6.2
Leak in the blood
pressure cuff or air hose
Replace the blood pressure cuff with the air hose.