Manufacturer’s EMC guidelines
48
Measurement of
interference emissions
IEC 60601-1 test
level
Compliance level
Electromagnetic environment
–
guidelines
IEC 61000-4-4
Not applicable
IEC 61000-4-5
Not applicable
IEC 61000-4-6
Not applicable
Magnetic field at supply
frequency (50/60 Hz) in
accordance with
IEC 61000-4-8
30 A/m
30 A/m
Magnet fields at mains frequency should
match the typical values found in
business and hospital environments.
IEC 61000-4-11
Not applicable