Service Manual ToFscan - Rev G
8
3 battery levels
Green:
good
Yellow:
medium
Red:
low
Battery charging/mains power supply
or
50 mA
Stimulation Intensity:
(between 20 and 60mA, default value 50mA)
Software and mother board Version
The software version can be checked in the “parameter” menu top right of the display:
ToFscan software version 1.7.x and below
ToFscan software version 1.8.x
The version has the following format:
1.X.Y A
or
1.X.Y C
X:
will change with new functionalities
Y:
will change with technical release
A or C:
change with main electronic board version. C version is for board 2.51 and higher, A is for
previous board (2.3c and 2.3d). This may change the way of assembling/disassembling and some
spare parts. See corresponding chapter.
Main unit diagnostic
The table below summarises a list of possible malfunctions and the solutions for resolving them.
Malfunction
Solution
Can be handled in
the first instance by
A. The device does not start
or stops on its own after a few
seconds
1. Set the device to charge
User
B. The device switches off as
soon as the power supply is
removed
1. Battery failure. Has to be changed
2. Charge circuitry failure. Mother electro-
nic board has to be checked
Distributor