Service Manual ToFscan - Rev G
23
Power consumption
• With power supply connected: 1 Watt while in use and 0,1 Watt while in sleep mode.
Environment
Shipping and storage conditions
The ToFscan and its accessories must be stored or transported under the following
restrictions and conditions. These conditions apply to non-operational storage and transport
conditions.
Temperature
10°C to +50°C
Humidity
15% to 95% (without condensation)
Pressure
500 hPa to 1060 hPa
For storage and transport, the original packaging must be used.
Protect the ToFscan from pikes in temperature which can cause condensation.
Operating environment
Reminders:
Risk of explosion: do not use the ToFscan in a flammable environment or in locations where
flammable anaesthetics may be accumulated.
The ToFscan is not designed to operate in an environment where there are SCANNERS, M.R.I.’s or any
other devices creating large magnetic fields. In order to limit electrostatic discharge, the humidity
must be maintaining above 35% and an antistatic floor is recommended.
The ToFscan is designed to operate safely under the following conditions. Situations other than those
described are likely to compromise the reliability of the device.
Temperature
10°C to +40°C
Humidity
35% to 90% (without condensation)
Pressure
700 hPa to 1060 hPa