Service Manual ToFscan - Rev G
20
with a Quaternary Ammonium Compound (QACs), isopropyl alcohol 70%. Before using any of these
solutions, refer to the manufacturer’s documentation and test on a reduced surface.
Example of recommended Quaternary Ammonium product:
- mikrozid® sensitive liquid from the manufacturer Schülke & Mayr GmbH.
Please check with your local authorized distributor or with the manufacturer which products are
available and approved in your country.
The ToFscan cable (electrode and/or sensor) must not come
into direct contact with, be
immersed in, splashed or filled with liquid and is to be cleaned in the same manner as the
ToFscan.
When cleaning the cables of the ToFscan, be careful not to create excessive traction on the splint that
could cause premature breakage of the wires inside the sheath.
End-of-life disposal / Recycling
In the interest of environmental safety, you are required to pass your used system on to a
collection body with the capability to treat devices containing electronic components and
Lithium-ion storage batteries.
To dispose of or recycle device components, contact a company specialising in the recycling of
electronic devices.
Unsorted electronic waste products are potentially hazardous to the environment.
Packaging materials must be disposed of or recycled in accordance with the regulations in force.
Technical specifications
The ToFscan comes with a microcontroller and a colour LCD screen for optimal legibility and ease
of use.
Safety
• Biocompatible material sensors (part in contact with the patient). Latex free
• Compliant with European directive CEE 93/42. Class 2a device (CE 0459 LNE/G-MED)
• Compliant with standards IEC 60601-1. Class II equipment.
• Compliant with standards IEC 60601-2-10.
• EMC: IEC 60601-1-2
EMC Emission
Emission test
Compliance
EMC Instructions/cautions
RF Emissions CISPR 11
Group 1
The ToFscan uses RF energy only for internal
functions. Therefore RF emissions are very low and
should not disturb other nearby devices.
RF Emissions CISPR 11
Harmonics IEC 61000-3-2
Class B
The ToFscan must be use in professional healthcare
facility environment
Voltage fluctuations
Class A
and flicker IEC 61000-3-3
Compliant
The ToFscan can be connected to the public mains
network