7. 

Check the pad attached to the underside of the monitor mount.  If there is any 

damage to the pad or if the pad is not properly attached to the unit, it must be replaced.

 

The procedures mentioned in these instructions are based on the recommendations of 
HHS, I.E.C., etc. 
 
In  cases  of  increased  wear  or  where  severe  working  conditions  exist,  maintenance 
checks should be made at shorter than the specified intervals. 
 
Should any control or indicator fail to operate properly, do not use the equipment until 
a  repair  has  been  affected.    Operating  equipment  with  defective  components  may 
expose the operator or the patient to safety hazards. 
 
The manufacturer strongly recommends that a maintenance program be initiated and 
that  a  maintenance  record  be  kept  detailing  dates  and  nature  of  maintenance 
performed, the name of the service engineer, and any other relevant information. 

 

WARNING! 

Removal  of  any  hardware  requires 
re-assembly  with  Loctite  #242  or 
equivalent (depending on hardware 
size) 

following 

manufacturer's 

instructions. 

 

Pad 

Summary of Contents for 1000F Series

Page 1: ...Image Diagnostics Inc Page 1 idi Manufacturer of Quality Medical Products Image Diagnostics Inc IDI 1000F Series Monitor Suspension System Model IDI 1000F 1 thru 1000F 6 With OTC Adapter Installation...

Page 2: ...of this manual was originally written approved and published by the manufacturer in English The information in this manual is subject to change without notice CONTENTS SECTION TITLE A INTRODUCTION PRO...

Page 3: ...el mounting and positioning The mobile package includes clip mounted bridge rails PRODUCT DATA Package Includes Mobile gantry and counterbalanced swing arm with 42 sweep radius 25 vertical travel and...

Page 4: ...g documents Failure to follow these instructions can cause accidents resulting in serious injury to patient user and damage to equipment Caution risk of electrical shock Protective Ground This is the...

Page 5: ...with state federal and local recommendations concerning electrical safety when used in medical installations All electrical connections shall be done by licensed approved electrician per national ele...

Page 6: ...upport Image Diagnostics Inc 310 Authority Drive FITCHBURG MA 01420 USA Or call IDI at 978 829 0009 or send fax to 978 829 0027 Or toll free 877 304 5434 SAFETY HAZARDS ALERTS Alert Circumstances for...

Page 7: ...em 9 and two rail spacers item 14 up to the adapter carriages item 1 and hold in place Place a pair of rail clips item 15 around the rail flanges locating rail clips under the mounting holes in the ad...

Page 8: ...our assemblies in until the wheels contact the OTC rails then tighten the nut plate screws Use Loctite 242 on hardware INSTALL RAIL END STOP Refer to End Stop Assembly Drawing NOTE End plates have bee...

Page 9: ...allet up and into the rails Adjust center wheels if necessary per figure on page 8 Make sure all carriage wheels are engaged onto rails before lowering Hi Jack Remove pallet and shipping bars from sup...

Page 10: ...s in until they contact the rotation cap item 43 The setscrews may be tightened or loosened later to adjust rotational resistance WARNING Failure to securely attach hardware may result in serious inju...

Page 11: ...l strip item 31 has been provided for distribution of line voltage to each monitor Cut and strip each monitor power cord to fit WARNING Remove plastic cover and remove block of wood between gantry cor...

Page 12: ...nection shall have double insulation Conduit provided for routing cables shall not be relied on for insulation 5 Connect all monitor ground wires and the service ground wire as shown in the figure abo...

Page 13: ...nut plates item 4 of two hanger standoff and safety bracket assemblies into rail track After applying LOCTITE 242 to the end that will be secured into the nutplate position standoffs approximately 48...

Page 14: ...pounds 4 2 6 0 Nm Use LOCTITE 242 on hardware CAUTION Over tightening the standoff will weaken or break the set screw connecting the standoff to the nut plate causing attached components to fall DO NO...

Page 15: ...d is to be attached to pivot tube and to be routed underneath the arm item 51 via the cable clamps items 46 50 Rotate entire system and bring to maximum positions the system should not drift from any...

Page 16: ...izontal Fine adjustment to compensate monitor counterbalance can be made by turning adjustment screw item 53 Rotate adjustment nut clockwise to increase capacity counter clockwise to decrease capacity...

Page 17: ...s with a clean cloth dampened with disinfectant or a mild detergent solution Do not use abrasives solvents sprays or corrosive cleaning agents Gently rub with a clean soft cloth to dry If room is to b...

Page 18: ...sign reading DO NOT USE so that the system is not used by mistake Then contact your Toshiba representative for inspection and repair 2 Visual check Before checking the operation confirm the following...

Page 19: ...ion does not lack smoothness compared to when the monitor suspension was installed There is no play There are no abnormal sounds The monitors or monitor suspension are not tilted due to looseness of s...

Page 20: ...ete range Raise and lower monitor mount assembly Roll carriages through complete range of rails 5 Clean and touch up the painted surfaces 6 Check that the hardware is tight and that there is no wear a...

Page 21: ...st maintenance checks should be made at shorter than the specified intervals Should any control or indicator fail to operate properly do not use the equipment until a repair has been affected Operatin...

Page 22: ...ly recycled The gas springs are nearly 100 recyclable They must be decompressed with the remaining oil drained before processing The compressed gas inside is nitrogen J APPROVALS SGS Testing Service T...

Page 23: ...Installation and Maintenance Guide for IDI 1000F Series Suspension System for use with OTC Adapter D000 571 REV E MAY 2018 Image Diagnostics Inc Page 23 E May 2018 ECO E571 Added warnings for pad...

Reviews: