Installation and Maintenance Guide for IDI 1000F Series Suspension System for use
with OTC Adapter
D000-571 REV E
MAY-2018
Image Diagnostics, Inc.
Page
10
3.
Take the thicker thrust ring (item 20) and put around rotation bearing (item 50).
4.
Place the stainless-steel thrust ring (item 47) below the thicker thrust ring (item
20).
5.
The six bolts (item 16) with lockwashers (item 29) go through rotation cap (item
43) and rotation bearing (item 50) into the pivot tube (item 43 in the Core
Assembly Drawing). Use
LOCTITE #242
on hardware.
6.
Apply a very little amount of
LOCTITE #242
to and insert the six setscrews
(item 24) into the six threaded holes on second rotation cap (item 43). Thread
the setscrews in until they contact the rotation cap (item 43). The setscrews
may be tightened or loosened later to adjust rotational resistance.
WARNING!
Failure to securely attach hardware
may result in serious injury to
patient, user, and damage to
equipment.
INSTALL MONITOR (Refer to Monitor Mount Assembly Drawing)
Maximum weight of any individual monitor.
Maximum total weight of monitors.
1.
To attach a monitor to the system, first remove a monitor mounting bracket
(item #30) by removing its mounting screw (item #21). Fasten the bracket to
the monitor (hardware not provided). Reinstall the bracket with the monitor.
WARNING!
Failure to securely attach hardware
may result in serious injury to
patient, user, and damage to
equipment.
2.
Monitors can be tilted and rotated.
3.
If monitor tilt resistance is inadequate and it fails to stay in the desired position,
tighten bolt (item 21).
25 LBS
11 kg
44 kg
96 LBS