Daily Maintenance Checks:
1.
Pre-Operational and Post-Operational Checks
Perform daily checks of the monitor suspension BEFORE and AFTER operating
the equipment.
CAUTION! If any abnormality is found in the monitor
suspension, stop using it. Post a sign reading
"DO NOT USE" so that the system is not used
by mistake. Then contact your Toshiba
representative for inspection and repair.
2.
Visual check
Before checking the operation, confirm the following:
•
The monitor suspension is not tilted. Refer to the following figures.
•
Although "A" in the following figure shows a gap, there is no gap at the
monitor frame rotation section.
•
Monitors installed on the monitor suspension are not tilted.
•
Screws or some parts are not loosened or removed.
•
The pad mounted to the bottom of the monitor mount assembly is not
damaged or missing.
TILTED STATUS
NORMAL STATUS
A