![IDEAL INDUSTRIES 61-484 Service Information Download Page 22](http://html.mh-extra.com/html/ideal-industries/61-484/61-484_service-information_617128022.webp)
Page 20
Calibration Adjustment Point
AC and DC Volts Table
Function
Range
Standard Value
61-486
61-484
DC mV
60.00mV
30.00mV
O
55.00mV
O
600.0mV
300.0mV
O
O
550.0mV
O
O
DCV
6.000V
3.000V
O
O
5.500V
O
O
60.00V
30.00V
O
O
55.00V
O
O
600.0V
300.0V
O
O
550.0V
O
O
1000V
950V
O
O
AC mV
60.00mV
2.00mV, 50Hz
O
10.00mV, 50Hz
O
55.00mV, 50Hz
O
600.0mV
20.0mV, 50Hz
O
O
100.0mV, 50Hz
O
O
550.0mV, 50Hz
O
O
ACV
6.000V
0.200V, 60Hz
O
O
1.000V, 60Hz
O
O
5.000V, 60Hz
O
O
60.00V
2.00V, 60Hz
O
O
10.00V, 60Hz
O
O
50.00V, 60Hz
O
O
600.0V
20.0V, 60Hz
O
O
100.0V, 60Hz
O
O
500.0V, 60Hz
O
O
1000V
200V, 60Hz
O
O
500V, 60Hz
O
O
950V, 60Hz
O
O