![IDEAL INDUSTRIES 61-484 Service Information Download Page 18](http://html.mh-extra.com/html/ideal-industries/61-484/61-484_service-information_617128018.webp)
Page 16
Table 3 Capacitance Test:
Step
Input
Reading
1
0.900µF
0.887µF
~
0.913µF
2
9.00µF
8.87µF
~
9.13µF
3
90.0µF
88.7µF
~
91.3µF
4
0.900mF
0.887mF
~
0.913mF
5
9.00mF
8.87mF
~
9.13mF
Testing the Diode Function
To verify accuracy in the AC and DC capacitance function, do the following:
1.
Turn on the meter, and turn the rotary switch to µF position.
2.
Connect the Calibrator to the
VΩ
and
COM
inputs on the Meter.
3.
Set the Calibrator for the voltage from step 1 in Table 4.
4.
Compare the reading on the Meter display with the display reading shown in Table 4.
5.
Connect a silicon diode to the
VΩ
and
COM
inputs on the Meter.
6.
Compare the reading on the Meter display with the display reading shown in Table 4.
7.
If the display reading falls outside of the range shown in Table 4, the Meter does not meet
specification.
Table 4 Diode Test:
Step
Input
Reading
1
1.800V
1.771V
~
1.829V
2
Silicon Diode
0.450V
~
0.650V
Testing the mA Function
To verify accuracy in the AC and DC mA ranges, do the following:
1.
Turn on the meter, and turn the rotary switch to the correct position.
2.
Connect the Calibrator to the
mA
and
COM
inputs on the Meter.
3.
Set the Calibrator for the voltage and frequency from step 1 to 6 in Table 5.
4.
Compare the reading on the Meter display with the reading shown in Table 5.
5.
If the display reading falls outside of the range shown in Table 5, the Meter does not meet
specification.