![IDEAL INDUSTRIES 61-484 Service Information Download Page 19](http://html.mh-extra.com/html/ideal-industries/61-484/61-484_service-information_617128019.webp)
Page 17
Table 5 mA Test:
Step
Input
Frequency
Reading
1
50.00mA AC
50Hz
49.37mA
~
50.63mA
2
50.00mA AC
1000Hz
49.37mA
~
50.63mA
3
500.0mA AC
50Hz
493.7mA
~
506.3mA
4
500.0mA AC
1000Hz
493.7mA
~
506.3mA
5
50.00mA DC
49.57mA
~
50.43mA
6
500.0mA DC
495.7mA
~
504.3mA
Testing the Ampere Function
To verify accuracy in the AC and DC ampere ranges, do the following:
1.
Turn on the meter, and turn the rotary switch to the correct position.
2.
Connect the Calibrator to the
A
and
COM
inputs on the Meter.
3.
Set the Calibrator for the voltage and frequency from step 1 to 20 in Table 6.
4.
Compare the reading on the Meter display with the reading shown in Table 6.
5.
If the display reading falls outside of the range shown in Table 6, the Meter does not meet
specification.
6.
Table 6 Ampere Test:
Step
Input
Frequency
Reading
1
5.000A AC
50Hz
4.937A
~
5.063A
2
5.000A AC
1000Hz
4.937A
~
5.063A
3
9.00A AC
50Hz
8.86A
~
9.14A
4
9.00A AC
1000Hz
8.86A
~
9.14A
5
5.000A DC
4.957A
~
5.043A
6
9.00A DC
8.90A
~
9.10A
Testing the Temperature Function
To verify accuracy in the temperature function, do the following:
1.
Turn on the meter, and turn the rotary switch to the temperature position.
2.
Connect the Calibrator to the
VΩ
and
COM
inputs on the Meter.
3.
Set the Calibrator for the temperature from step 1 to 5 in Table 7.
4.
Compare the reading on the Meter display with the reading shown in Table 7.
5.
If the display reading falls outside of the range shown in Table 7, the Meter does not meet
specification.