id LAB FOTOMETER 2008 User Manual Download Page 6

 

Setting reference 

To set reference press SET button bellow the reference you want to set. Current intensity 
value is stored as reference value.  

Using input filter 

Light detector has two input filters. Fast filter is intended for measurement of values 
changeable in time. It provides fast response with slightly increased noise level. Slow filter 
suppress noise but has rather slow response time. Thus it is suitable for measurement of in 
stable conditions when light intensity does not change fast. 

Setting input filter 

To adjust input filter press the 

FILTER

 button. The filter will toggle from slow to fast and 

vice versa with every pres of the button. 

Diagnostic modes 

The device provides four diagnostic modes which are useful for troubleshooting. The 
diagnostic modes are: zero offset, full scale, negative full scale, common mode rejection.  
 

Mode 

Description 

Zero offset 

Both differential inputs of A/D converter are switched to ground. 

Full scale 

Negative input of A/D converter is grounded. Positive is switched 
to the voltage reference. 

Negative full scale 

Positive input of A/D converter is grounded. Negative is switched 
to the voltage reference. 

Common mode rejection  Both differential inputs of A/D converter are switched to voltage 

reference. 

Switching to diagnostic mode 

To switch to diagnostic mode press 

DIAG

 button. First diagnostic mode is selected. With 

every consecutive press of

 DIAG

 button device will switch to next diagnostic mode. After 

last diagnostic mode the device switch back to normal measurement mode. 
While in diagnostic mode you can switch to normal measurement mode anytime by pressing 

RANGE+

RANGE-

 or 

AUTO/MANUAL

 buttons. 

Basic maintenance tasks 

Replacing bulb in the light source 

To replace the bulb in the light source follow the steps described bellow. 

 

Switch off the device. 

 

Disconnect cable between control unit and the light source. 

 

Remove four screws on the back cover of the light source using Philips screwdriver. 

 

Carefully remove the back cover together with cylindrical side cover. 

 

Disconnect internal connector so you can completely remove back and side cover. 

 

Remove four screws which hold the bulb assembly and remove bulb assembly. 

 

Loosen screws holding the bulb in place and remove old bulb. 

 

Install new bulb in the same position as the original and tighten screws. 

 

Put bulb assembly back to its place and fasten it by four screws. 