id LAB FOTOMETER 2008 User Manual Download Page 4

 

Display content 

Display on control unit shows measured light intensity value as well as other related 
information like actual range, measurement mode, selected input filter and so on. The layout 
of the display is described in the table below. 
 
Switching status 

 

 

Mode 

Overflow status 

 

 

Range number 

Intensity value 

Reference 1 

Reference 2 

Diagnostic mode 

Range 

[dB] 

[dB] 

Filter 

 
Meaning of display messages is described bellow. 
 

Display field 

Possible values 

Description 

SWITCHING 

Device is switching between ranges. 
Intensity value is not valid at the moment. 

Switching status 

<blank> 

Device is not switching range. 

OVERFLOW 

Light detector is saturated by ambient light 
or too strong signal from light source. . 
Intensity value is not valid at the moment. 

Overflow status 

<blank> 

Light detector is not in overflow condition. 

Intensity value 

<value in AU> 

Actual value of intensity in AU. Selected 
range is already taken into account. 

Range 100 

Selected range 0..100 AU (most sensitive) 

Range 1000 

Selected range 0..1000 AU 

Range 10000 

Selected range 0..10000 AU 

Range  

Range 100000 

Selected range 0..100000 AU (less sensitive) 

Reference 1 

<value in dB> 

Relative change in intensity since the 
reference 1 was set expressed in dB.  

Reference 2 

<value in dB> 

Relative change in intensity since the 
reference 2 was set expressed in dB.  

AUTO 

Device is in automatic mode. 

Mode 

MANUAL 

Device is in manual mode. 

RANGE 0 

Selected range 0 (0..100 AU)  

RANGE 1 

Selected range 1 (0..1000 AU) 

RANGE 2 

Selected range 2 (0..10000 AU) 

Range number 
(according to codes used 
when operating device 
from remote computer) 

RANGE 3 

Selected range 3 (0..100000 AU) 

<blank> 

Device is in standard measurement mode. 

DIAG ZERO 

Device is in diagnostic mode ZERO. 

DIAG FULL 

Device is in diagnostic mode FULL SCALE. 

DIAG NEG 

Device is in diagnostic mode NEGATIVE 
FULL SCALE. 

Diagnostic mode 

DIAG CMR 

Device is in diagnostic mode COMMON 
MODE REJECTION. 

FILTER SLOW  Slow input filter selected. 

Filter 

FILTER FAST 

Fast input filter selected. 