id LAB FOTOMETER 2008 User Manual Download Page 5

 

Operating device in automatic mode 

In automatic mode the device switch to more sensitive range, when intensity value decrease 
below 8 percent of current range. When intensity rise over 100 percent of current range, 
device automatically switches to less sensitive range. 
 

During switching device displays SWITCHING message on the display. During this 

period the intensity readout is not valid. On some conditions (slow filter selected, large 
sudden change in intensity) the intensity might not be settled when SWITCHING message 
disappear and additional waiting for valid value might be necessary. 

Switching to automatic mode 

While in manual mode, press 

AUTO/MANUAL

 button to switch to automatic mode. 

Switching back to manual mode 

While in automatic mode you can press 

AUTO/MANUAL

 button to switch back to manual 

mode. By pressing 

RANGE+

RANGE-

 or 

DIAG

 while in automatic mode you switch back 

to manual mode as well. 

Operating device in manual mode 

In manual mode, ranges must be switched manually by the user. When range is manually 
switched, device displays SWITCHING message and goes through the switching procedure in 
same manner as in automatic mode. 

Switching to manual mode 

While in automatic mode you can press 

AUTO/MANUAL

 button to switch to manual mode. 

By pressing 

RANGE+

RANGE-

 or 

DIAG

 while in automatic mode you switch to manual 

mode as well. 

Switching back to automatic mode 

While in manual mode, press 

AUTO/MANUAL

 button to switch to automatic mode. 

Switching to less sensitive range 

To switch to less sensitive range press 

RANGE-

 button. 

Switching to more sensitive range 

To switch to more sensitive range press 

RANGE+

 button. 

Using references 

To compute light attenuation directly during measurement there are two user references 
available. When reference is set it displays the ratio between reference value and current value 
in a logarithmic scale in dB. The value in logarithmic scale is computed using following 
formula: 
 





=

]

[

Re

]

[

log

10

]

[

AU

value

ference

AU

value

Intensity

dB

value

Log

 