id LAB FOTOMETER 2008 User Manual Download Page 3

 

Device description 

Light attenuation meter FOTOMETER 2008 consists of modulated light source, lock-

in type light detector and control unit. Device measures intensity of light traveling from the 
light source to the light detector. It makes the device suitable for measurement changes of 
attenuation of light in the area between the light source and the light detector. Due to 
synchronization between light source modulation and the light detector the device is, to some 
extent, immune to ambient light and could be operated on normal day light conditions. 

The readout of intensity value is in linear arbitrary units (AU) and is not directly 

related to any engineering units. The device is intended only for measurement of relative 
changes in light intensity.  
 

Device has four decadic ranges and could be operated in two different modes. In 

manual mode

 the ranges must be switched manually by user. In 

automatic mode

 the device 

itself changes mode when intensity readout goes out of actual range. 
 

The light detector is equipped with two input filters. 

Fast filter

 provides quick 

response but slightly higher noise and is intended for measurement of unstable, changing light 
intensity. For measurement in stable, balanced conditions might be done using 

slow filter

 

which reduces noise but has very long settling time (up to one minute). 

Safety rules 

The control unit is powered by 230V AC which might be dangerous to life. Never remove the 
cover of control unit when the mains electricity is connected. Mains plug MUST be always 
removed before unscrewing the cover. Any service must be provided by a qualified electrician 
only. 
Device may be operated only in room environment between 15 and 40 C with non condensing 
humidity. When moving device from cold environment to room temperature always allow the 
device temperature to accommodate to room environment to avoid moisture to condensate on 
the device before switching the device on. 

Setting up the device 

Mount control device to its place, preferably a 19 inch rack case. Connect 15 pin cable 
between the control unit and the detector and 25 pin cable between the control unit and the 
light source. Double check that cables are connected properly according to labels on control 
unit. Connect mains electricity and switch the device on. 

The light source and light detector must be mounted in such position that light source 

outlet is facing the light detector inlet directly. The light beam must go through the area where 
the light attenuation should be measured. Make sure that axes of light source outlet and light 
detector inlet are perfectly parallel. Using small piece of paper make sure that the light beam 
is aimed directly to the center of light detector inlet. 