Firmware Version 1.2
MultiSystem 4070
117
Special functions
DEU
The linearization table can be utilized to compensate for sensor non-lineari-
ties. By calibrating a sensor, you will obtain this table, which can be entered
into the measuring instrument. Five different linearization tables, each with ten
value pairs, are available for each measurement channel.
1
Select the option
YES
at the
Linearisation
menu option.
.
2
Highlight
Table
:
.
3
Either highlight a stored table, or an empty line if you want to enter a new
table:
.
4
Highlight the
Name
entry:
.
5
Enter a name for the new table:
.
6
Highlight
Reference point 1
:
.
7
Enter the first set value:
.
8
Enter the first actual value:
.
9
Repeat steps 7 and 8 for all required lines of the table.
10
Complete the entry of set and actual values:
.
11
Confirm changes and exit dialog:
The new table is selected as active.