PEX/PIO/PISO-DA Series Card
Analog Output Boards
User Manual/ Ver. 3.1/ Oct. 2013/ PMH-0010-31/ Page: 32
The steps required to calibrate the current for channel_n are as follows:
Step 1:
DaValue=0x1fff
Step 2:
Send the DaValue to hannel_n on the PIO/PISO card
Step 3:
Measure the current of channel_n on the I-7017
If this value is >= 0 mA, then go to Step 5
Step 4:
Increase the DaValue, the return to Step 2
Step 5:
MinI[n]=DaValue-1
Step 6:
DaValue=0x3fff
Step 7:
Send the DaValue to channel_n on the PIO/PISO card
Step 8:
Measure the current of channel_n on the I-7017
If this value is >= 20 mA, than go to Step 10
Step 9:
Increase the DaValue, the return to Step 7
Step 10:
MaxI[n]=DaValue