![background image](http://html2.mh-extra.com/html/icp-das-usa/pcie-8620/pcie-8620_user-manual_4949528084.webp)
High-speed Multifunction Boards
User Manual, Ver. 1.0, May 2015, PMH-029-10 Page 84
Step 5:
Calibrate the Analog Input Channel 3
(1)
Connect 9.9 V voltage source to PCIe-8620.AI3 (Pin4)
(2)
Connect GND source to PCIe-8620.A_GND (Pin14)
(3)
Click
“
Read
”
button to get hexadecimal value
(4)
Connect 0 V voltage source to PCIe-8620.AI3 (Pin4)
(5)
Connect GND source to PCIe-8620.A_GND (Pin14)
(6)
Click
“
Read
”
button to get hexadecimal value
(7)
Connect 4.9 V voltage source to PCIe-8620.AI3 (Pin4)
(8)
Connect GND source to PCIe-8620.A_GND (Pin14)
(9)
Click
“
Read
”
button to get hexadecimal value
(10)
Connect 0 V voltage source to PCIe-8620.AI3 (Pin4)
(11)
Connect GND source to PCIe-8620.A_GND (Pin14)
(12)
Click
“
Read
”
button to get hexadecimal value