D300 Series Operating Manual
23
Please enter the pH value obtained from the standard portable instrument or standard laboratory test equipment data. After pressing
‘Enter’, the instrument will display the calibration result, press the “Save” button to exit:
Save the results and it can be queried next time 2.4 ORP Standard solution calibration (S-S-C) Choose ORP single point calibration, and put
the sensor into standard liquid, follow the instruction to operate:
Save the results and it can be queried next time 2.5 ORP onsite calibration Choose onsite ORP and put sensor into water sample,follow the
instruction to operate;
Input the ORP value obtained from the standard portable instrument or standard laboratory laboratory equipment data. After pressing
[ENT] the instrument will display the calibration result, press the “Save” button to exit;
Save the results and it can be queried next time.