D300 Series Operating Manual
21
Choose ‘pH (2 points) 4/6/9’ (Before each operation, the electrode can be placed in the standard solution for about 60s)
After calibration of 1st point, follow with the 2nd point:
After above calibration, there will be the results:
Calibration 6.86 and 4.00 will be saved in offset 1 and slope 1; calibration 6.86 and 9.18 will be saved in offset 2 and slope 2.
Save the results and it can be queried next time.
The pH value of the buffer is measured at 25 °C, please select the correct buffer
(please refer to P30 “Buffer Table”).
Before calibration, the electrode can be activated by soaking in the solution to ensure
the stability and accuracy of the calibration value and the detection value.
Calibration with Standard Liquid (pH, 3 points)
There are two groups of 3-point standard liquid USA: 4.01, 7.01, 10.01 and NIST:
4.00, 6.86, 9.18, The following example uses the NIST standard solution set:
Choose pH standard liquid, press
[ENT] to enter the mode
Summary of Contents for ProCon D300 Series
Page 1: ...D300 Series pH ORP Digital Analyzer Controller Operating Manual ...
Page 7: ...D300 Series Operating Manual 7 Keypad Descriptions ...
Page 12: ...D300 Series Operating Manual 12 Measurement Settings Quick Start ...
Page 13: ...D300 Series Operating Manual 13 Calibration Settings Quick Start ...
Page 14: ...D300 Series Operating Manual 14 Alarm Settings Quick Start ...
Page 15: ...D300 Series Operating Manual 15 Current Settings Quick Start ...
Page 16: ...D300 Series Operating Manual 16 Communication Settings Quick Start ...
Page 17: ...D300 Series Operating Manual 17 System Settings Quick Start ...
Page 18: ...D300 Series Operating Manual 18 Maintenance Settings Quick Start ...