SERVICE   PROCEDURES   

   

 

Replace   the   batteries   when   the   <bAtt>   message   appears   (page   7).   

 

Change   or   clean   the   probe   base   if   the   probe   does   not   move   smoothly   (page   6).   

 

The   device   can   be   cleaned   with   a   damp   cloth   containing   disinfectant   according   the   instructions   (page   6).   

 

There   are   no   other   service   procedures   that   can   be   carried   out   by   the   user.   All   other   servicing   and   repairs   must   be   carried   out   by   the   
manufacturer   or   certified   service   facilities.   
   

   

DIAGRAM   OF   TONOMETER   FUNCTIONS   

   

 

   

Summary of Contents for Tonovet TV01

Page 1: ...www icaretonometer com 1 ENGLISH USER S AND MAINTENANCE MANUAL...

Page 2: ...ning the probe base 6 Cleaning the tonometer surface 6 Replacing the batteries 7 Error messages 7 Trouble shooting 7 Service procedures 8 Diagram of tonometer functions 8 Technical information 9 Symbo...

Page 3: ...he substances that are flammable including flammable anaesthetic agents Check that a clean new disposable probe is used before each measurement To avoid transmitting certain microbiological agents e g...

Page 4: ...EASURING THE INTRAOCULAR PRESSURE The tonometer needs to be in a horizontal position during the measurement The distance should be 4 8 mm 1 6 1 3 inch from the probe to the cornea TIP The distance fro...

Page 5: ...1 8 2 5 Slight deviation the effect to the result is not likely to be relevant d Line in the middle the letter blinks 2 5 3 5 Deviation is greater than normal but the effect on the result is usually n...

Page 6: ...ase every 12 months or if the cleaning is not effective CLEANING THE PROBE BASE The probe base should be cleaned every 6 months The probe base can be reused after careful cleaning Make sure the probe...

Page 7: ...h a new one If the error message continues to appear replace the probe base E 02 The probe did not touch the eye The measurement was made from too far away or the probe base is dirty E 03 The probe sp...

Page 8: ...e probe base if the probe does not move smoothly page 6 The device can be cleaned with a damp cloth containing disinfectant according the instructions page 6 There are no other service procedures that...

Page 9: ...cal shock protection Operation environment o Temperature 10 C to 35 C o Relative humidity 30 to 90 o Atmospheric pressure 800 hPa 1060 hPa Storage environment o Temperature 10 C to 55 C o Relative hum...

Page 10: ...Icare TONOVET tonometer TV01 manual English www icaretonometer com 10...

Reviews: